Table 8 Menu option descriptions (continued)
Menu
Description
Settings
Sets the time zone and the network settings. Shows the analyzer information.
Power
Shuts down the analyzer. Prepares the analyzer to be disconnected from the electrical outlet.
Operation
N O T I C E
Hach recommends that the analyzer is set to idle mode at the end of each treatment day and set back to
measurement mode at the start of each treatment day. This will make sure that the analyzer completes the
necessary self-diagnostic tests at the start of each treatment day, such as the performance check of the remote
indicator lights and speaker.
Set to measurement mode
1.
When "Idle" shows on the home screen, push
idle
and select continue. The analyzer goes to
measurement mode.
The first measurement shows on the display in 10 minutes. Then a new measurement shows on
the display at intervals of 5 minutes
20
.
2.
Fully close the door to keep the enclosure rating.
Set to idle mode
N O T I C E
Do not keep the analyzer in idle mode continuously for more than 3 days or damage to the analyzer can occur. If
the analyzer will not be used for more than 3 days, prepare the analyzer for storage. Refer to
on page 62.
To set the analyzer to idle mode, push
idle
and select continue.
In idle mode, the status of the analyzer and remote indicator are as follows:
• The analyzer stays on.
• Chlorine measurements stops.
• Reagent use stops.
• Sample water use stops.
• The analyzer status lights on the remote indicator are on (flashing blue, stable amber or stable
red).
• The chlorine status lights on the remote indicator are off.
Show the analyzer information
1.
Push
menu
.
2.
Select Settings.
3.
Push the
DOWN
arrow to until "About" is selected. Push
ok
.
The analyzer information shows.
20
Measurements may be aborted when the system detects a measurement quality issue. The
maximum possible measurement interval is 20 minutes, after such time the unit will alarm and go
into idle mode.
English
55
Содержание CM130
Страница 1: ...DOC023 53 80540 CM130 11 2018 Edition 5 User Manual...
Страница 2: ......
Страница 18: ...Figure 9 Analyzer dimensions 16 English...
Страница 65: ...Figure 22 Remove the pump clamp English 63...
Страница 77: ...English 75...
Страница 79: ...English 77...
Страница 80: ...78 English...
Страница 81: ...English 79...
Страница 82: ...80 English...
Страница 83: ...English 81...
Страница 84: ...82 English...
Страница 85: ...English 83...
Страница 86: ...84 English...
Страница 87: ...Appendix C Screens Data export This appendix shows the analyzer display screens for reference only English 85...
Страница 88: ...86 English...
Страница 90: ...88 English...
Страница 92: ...90 English...
Страница 94: ...92 English...
Страница 95: ...English 93...
Страница 96: ...94 English...
Страница 98: ...96 English...
Страница 99: ...English 97...
Страница 100: ...98 English...
Страница 101: ...English 99...
Страница 102: ...100 English...
Страница 103: ...English 101...
Страница 104: ...102 English...
Страница 105: ...English 103...
Страница 106: ...104 English...
Страница 107: ...English 105...
Страница 111: ......