3.
Push the black tubing of the grab sample assembly into the sample inlet fitting. The sample inlet
fitting is behind the metal colorimeter inlet fitting.
4.
Plumb the sample inlet tubing as follows:
a.
Plumb one end of the coil of black tubing to the sample source between the primary and
secondary carbon tanks, as identified in ANSI/AAMI/ISO 13959
Water for hemodialysis and
related therapies
.
b.
Hold the open end of the black tubing over a drain, then open the valve at the sample source.
Rinse the tubing for 30 seconds.
c.
Push the open end of the coil of black tubing into the check valve of the grab sample
assembly.
Decrease the length of the black tubing to remove the unnecessary length. Leave at least 5 feet
of length.
Note: AAMI is the acronym for Association for the Advancement of Medical Instrumentation.
22
English
Содержание CM130
Страница 1: ...DOC023 53 80540 CM130 11 2018 Edition 5 User Manual...
Страница 2: ......
Страница 18: ...Figure 9 Analyzer dimensions 16 English...
Страница 65: ...Figure 22 Remove the pump clamp English 63...
Страница 77: ...English 75...
Страница 79: ...English 77...
Страница 80: ...78 English...
Страница 81: ...English 79...
Страница 82: ...80 English...
Страница 83: ...English 81...
Страница 84: ...82 English...
Страница 85: ...English 83...
Страница 86: ...84 English...
Страница 87: ...Appendix C Screens Data export This appendix shows the analyzer display screens for reference only English 85...
Страница 88: ...86 English...
Страница 90: ...88 English...
Страница 92: ...90 English...
Страница 94: ...92 English...
Страница 95: ...English 93...
Страница 96: ...94 English...
Страница 98: ...96 English...
Страница 99: ...English 97...
Страница 100: ...98 English...
Страница 101: ...English 99...
Страница 102: ...100 English...
Страница 103: ...English 101...
Страница 104: ...102 English...
Страница 105: ...English 103...
Страница 106: ...104 English...
Страница 107: ...English 105...
Страница 111: ......