Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification
Details
Dimensions
(W x H x D)
Analyzer: 37.0 x 53.8 x 22.6 cm (14.6 x 21.2 x 8.9 in.)
Remote indicator: 12.3 x 29.4 x 10.7 cm (4.8 x 11.6 x 4.2 in.)
Enclosure
Enclosure rating: IP52 rated with the door closed
Enclosure material: PC/ABS case, PC door, PC hinges and latches, 316 SST
hardware
Remote indicator material: PC/ABS
Display
10.9 cm (4.3 in.) color display, back-lit LCD, WQVGA
Weight
Analyzer: 8.9 kg (19.6 lbs)
Remote indicator: 1.1 kg (2.5 lbs)
Mounting
Analyzer: Wall
Remote indicator: Wall or ceiling
Protection class
I
Pollution degree
2
Installation category
II
Power requirements
Analyzer: 100 to 240 VAC, 50/60 Hz; 1 A maximum
Remote indicator: Power is supplied by the analyzer.
Power-loss alarm: Power is supplied by a lithium coin battery CR2032 (3 V) that is in
the remote indicator.
Recommended operating
temperature
15 to 30 °C (60 to 85 °F)
Safe operating
temperature
5 to 40 °C (41 to 104 °F)
Operating humidity
5 to 90% non-condensing at 30 °C (85 °F) maximum
Storage temperature
-20 to 60 °C (-4 to 140 °F)
Fittings
Sample line: ¼-in. OD push-to-connect fitting for plastic tubing
Drain line: slip-on fitting for ½-in. ID soft plastic tubing
Cabling
1
Ethernet: 100 m (328 ft) maximum, Ethernet cable with an RJ45 connector
Remote indicator: 762 m (2500 ft) maximum, CAT5e/CAT6 type Ethernet cable with
four twisted pairs of 24 AWG solid copper wiring.
Make sure that the cable shielding includes an inner foil wrap around all of the pairs
and an outer braid, also known as SFTP or SF/UTP. Do not use stranded or copper-
clad aluminum conductors.
Sample
Pressure: 276 to 689 kPa (40 to 100 psi) nominal
Pressure spikes: 827 kPa (120 psi) or less
Flow rate: 250 mL/minute minimum
Temperature: 15 to 30 °C (60 to 85 °F)
Filtration: 15 µm or less
Refer to
on page 72 interfering substances.
1
Cabling is supplied by the user.
English
3
Содержание CM130
Страница 1: ...DOC023 53 80540 CM130 11 2018 Edition 5 User Manual...
Страница 2: ......
Страница 18: ...Figure 9 Analyzer dimensions 16 English...
Страница 65: ...Figure 22 Remove the pump clamp English 63...
Страница 77: ...English 75...
Страница 79: ...English 77...
Страница 80: ...78 English...
Страница 81: ...English 79...
Страница 82: ...80 English...
Страница 83: ...English 81...
Страница 84: ...82 English...
Страница 85: ...English 83...
Страница 86: ...84 English...
Страница 87: ...Appendix C Screens Data export This appendix shows the analyzer display screens for reference only English 85...
Страница 88: ...86 English...
Страница 90: ...88 English...
Страница 92: ...90 English...
Страница 94: ...92 English...
Страница 95: ...English 93...
Страница 96: ...94 English...
Страница 98: ...96 English...
Страница 99: ...English 97...
Страница 100: ...98 English...
Страница 101: ...English 99...
Страница 102: ...100 English...
Страница 103: ...English 101...
Страница 104: ...102 English...
Страница 105: ...English 103...
Страница 106: ...104 English...
Страница 107: ...English 105...
Страница 111: ......