Network Termination
51350-18 Hardware Installation.fm
Section 5
5.4.2 Network Termination at the AquaTrend Interface without SOM
The following instructions apply only to the AquaTrend Interface without SOM,
Cat. No. 51200.
Note: To avoid damaging the
instrument, remove power to the
instrument before removing the
front cover.
The network signals require a terminator for proper operation. As shipped
from the factory, the terminator is disabled on all AquaTrend Interfaces as
shown in
.
IMPORTANT NOTE:Failure to properly terminate the network may cause total
network failure. Enable the network terminator at only one device.
Follow the steps below to enable the terminator and complete the installation
of the stand alone AquaTrend Interface.
1.
Remove the front panel of the AquaTrend Interface and locate J6 on the
circuit board in the back housing of the device. J6 is a white, three-pin
connector. A jumper connected only to pin 2 should be present as shown
in
2.
To enable the terminator, lift the jumper off of pin 2 and move the jumper
so it connects pins 2 and 3 as shown in
Figure 25
Network Termination at the AquaTrend Interface Jumper Position Enabled
Front Enclosure
Back Enclosure
(Some components omitted for clarity.)
J6
J6
U12
J3 1
19
2
20
U11
U5
J2
J7
1
9
U1
U13
TB1 1
6
1
3
SW1
DS1
J6
U12
TB1 1
6
1
3