Page 69
SILKOSTAT 9210
6. Maintenance
and
exploitation
A periodical maintenance will allow to have a good running and an exact
analysis. Verify the level of the calibration solution and the reagents and
refill if it is necessary.
Reagent refills
The analyser monitors the reagents and calibration solution consumption.
Messages are displayed when the canister levels are low.
Messages Signification Relays
on
Reagent < 250ml
Low reagent level
(range < 10 days)
Warning (Re 11)
Cal. Sol < 300ml
Calibration low level
(range < 2 calibration)
Warning (Re 11)
Short reagent
Reagent canister
volume < 150ml
System (Re 12)
Short cal.
Calibration solution
canister volume
< 150ml
It is necessary to program the new canister volume at each refill for a
good detection of the short reagent.
•
Stop the instrument "STOP".
•
Refill the reagent canisters.
•
Press the "MENU" key.
•
Choose the "MAINTENANCE" menu.
•
Choose the "REAGENTS REFILL" menu.
•
Choose CANISTERS FULL to program all the canisters
volumes to 2 litres.
REAGENTS REFILL
ADJUST VOLUMES
TUBES PRIMING
CANISTERS FULL yes/no
Select
Esc
ADJUST VOLUMES :
the 4 canisters volumes are displayed and
can be modified.
TUBES PRIMING :
start the priming / renewal procedure of the
reagents in the tubes. "Esc" allows to return to the previous
screen or to stop the procedure.
CANISTERS FULL:
choose this option each time than canisters
are full. The volumes of the 4 canisters are automatically
programmed with 2 litres.
Содержание POLYMETRON 9210
Страница 1: ...Operator Manual POLYMETRON 9210 221 192 010 Revision J 20 11 2007...
Страница 2: ......
Страница 10: ......
Страница 11: ...Page 5 SILKOSTAT 9210 1 Presentation of the instrument Figure 1...
Страница 18: ......
Страница 28: ......
Страница 74: ......
Страница 80: ......
Страница 114: ......
Страница 115: ......
Страница 116: ......