Page 49
SILKOSTAT 9210
Measure / Photometer test sub menu
Pumps-others sub menu
PHOTOMETER TEST
Dark current : 0.008
Reference : 3.50
Led current : 9000
Measure : 3.10
Absorbance : 52.7
Bck
ref
meas
Esc
With this menu, technicians can verify the good running order of
the photometer.
The mixer is operating and the transfer valve is open.
Displayed parameters
Dark current :
Measurement value which corresponds to the
emitter LED.
Reference :
reference signal after LED current adjustment.
Led current :
current of the LED.
Measure :
Continuous photometric measure signal.
Absorbance :
Absorbance in milli-Abs.
Function key
Bck :
allows to begin the measure which corresponds to the
switching off of the LED.
Ref. : allows to adjust the led current and to measure the
reference voltage.
Mes :
starts the continuous photometric measure.
MAINTENANCE
Pump
01 >
OFF
Pump
02
OFF
Pump
03
OFF
Pump
04
OFF
Drain
OFF
Mixer
OFF
Heater
OFF
Select
On
Off
Esc
This screen allows to test the good running order of the pumps,
the mixer, the drain and the heater.
Utilize the "Select" key to control an element and press the "On"
key to operate it or "Off" key to stop it.
The following table shows the nominal flow rates for each pump :
Pumps 01
Reagent 1
1,5ml/mn
Pumps 02
Reagent 2
1,5ml/mn
Pumps 03
Reagent 3
1,5ml/mn
Pumps 04
Cal. solution
10l/h
Содержание POLYMETRON 9210
Страница 1: ...Operator Manual POLYMETRON 9210 221 192 010 Revision J 20 11 2007...
Страница 2: ......
Страница 10: ......
Страница 11: ...Page 5 SILKOSTAT 9210 1 Presentation of the instrument Figure 1...
Страница 18: ......
Страница 28: ......
Страница 74: ......
Страница 80: ......
Страница 114: ......
Страница 115: ......
Страница 116: ......