Page 40
SILKOSTAT 9210
Programming sub menu (automatic execution)
CALIBRATION
EXECUTION
PRIMARY
EXECUTION
MANUAL
PARAMETERS
HISTORIC
Select
Esc
The calibration of the zero and the slope can be carried out
automatically at adjustable frequencies.
•
Press "CALIBRATION" menu.
•
Sub-menu "PROGRAMMING" and validate by pressing
"Enter".
•
Choose "PROGRAMMING" sub-menu and validate by pressing
“Enter”.
PROGRAMMING
OFFSET INTERVAL : 200 h
CAL. SOL.: 010.0 ppb
AUTO. CAL. : Non
Select
Esc
This screen allows to program the zero calibration frequency, the
calibration concentration solution and to choose an automatic
calibration or manual calibration. The following table indicates the
zero calibration frequency controlled in terms of the minimum
concentration of the samples.
PROGRAMMING
Sample minimum concentration
Zero frequency control
< 5 ppb
24 hours
5 < x < 50 ppb
48 hours
> 50 ppb
> 72 hours