Page 20
SILKOSTAT 9210
User signals connections
This box is found on the back of the SILKOSTAT on the top (see
figure 6).
Unscrew the 4 screws of the cover to gain access to the terminals.
Note :
The cable glands mounted on the “local controller box” should be
perforated with a screwdriver before being used.
To ensure a good seal, the external diameter of cables should be
between 5 and 7 mm.
Figure 12
“Local controller” connections
+
+
1
+
+
+
+
+
FIX2
+
Vc
c
TD
O
TC
K
TM
S
TD
I
GND
Rel 12
Re
l 1
1
Vout 1
Vout 2
Vout 3
Vout 4
Vout 5
Vout 6
Vout 7
Vout 8
Vout 9
Vout 10
Vout 11
Vout 12
Iout 1
Iout 2
Iout 3
Iout 4
Iout 5
Iout 6
Iout 7
Iout 8
In 1
In 2
In
3
In
4
In
5
In
6
In
7
In
8
In
9
In
10
In
11
In
12
Pt 100
He
at
er
Mixe
r
Rel
10
Rel 9
Rel 8
Rel 7
Rel 6
Rel 5
Rel 4
Rel 3
Rel 2
Rel 1
Danger :
The relay outputs can only supply power in very low safety voltage
(30 VAC Or 42.4 Vdc maximum) and limited to 0,5 A.
!
Содержание POLYMETRON 9210
Страница 1: ...Operator Manual POLYMETRON 9210 221 192 010 Revision J 20 11 2007...
Страница 2: ......
Страница 10: ......
Страница 11: ...Page 5 SILKOSTAT 9210 1 Presentation of the instrument Figure 1...
Страница 18: ......
Страница 28: ......
Страница 74: ......
Страница 80: ......
Страница 114: ......
Страница 115: ......
Страница 116: ......