Page 76
SILKOSTAT 9210 - MANUEL D'UTILISATION
Possible solutions
Problem
Symptom
Solution
Analyser stopped
•
System alarm : Rectify the problem according to
the
displayed message (see System Alarm Messages).
•
Start up the analyser (START key).
1. START / STOP
Mains power cut
off
•
Connect the analyser to the mains power.
•
Verify the fuse.
Doesn’t run
•
Test the pump : in the MAINTENANCE / PUMP menu
start pumps one after one and verify than their flow is
about 1,5ml / mn (30 pulses).
•
Change the pump.
2. PUMP
Irregular
•
The analyser may have been stopped without tubes
cleaning. Send water in the pump during its work.
•
Change the pump.
Defective tube
Change the tube
3. TUBES
Clogged tube
Change the tube
Solenoid valves
leak
Remove, test and clean.
The solenoid
valves can’t open
Sticked collar, must be clean
4. SOLENOID
VALVES
The solenoid
valves can’t close
Locked spring ; necessary cleaning
Dirty
Disconnect the samples from the sample block.
5. CAL.
Diminution of the
flow
Verify and adjust the flow (10 l/h), change the sample tube if it
is clogged.