![Hach Ultra POLYMETRON 9210 Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/hach-ultra/polymetron-9210/polymetron-9210_operators-manual_3141707051.webp)
Page 45
SILKOSTAT 9210
Maintenance menu
Note :
When one of the maintenance procedure is released, the analyser
automatically stops
Don’t forget to restart the instrument when maintenance
operations are realized.
Start up sub menu
MAINTENANCE
START UP
REAGENTS REFILL
EXTENDED STOP
TUBES CLEANING
PHOTOMETER TEST
PUMPS / OTHERS
SOLENOID VALVES
RELAYS
LOGICAL INPUTS
Select
Esc
This screen allows to choose the proposed maintenance help
procedure :
Start up
•
Starting
procedure.
Reagents refill
•
The procedure is applicable to each reagent refill.
Extended stop
•
Allows to condition the instrument before an extended stop.
Tubes cleaning
•
In case of reagent tubings clogging.
Photometer test : Check the photometer.
Pumps, Solenoid valves, relays, others
•
Working test of this devices.
Logical inputs
•
Logical inputs test.
START UP
Adjust flow
Sample
X
OK
Esc
This screen allows to adjust the sample flows
Adjust the first sample flow with the help of the screws placed in
the front panel of the face (see figure 3.4 item. 2) to have a
perpetual overflowing in the top of the overflow vessel.
Press OK to go to the next sample.
Warning :
Do not unscrew completely the flow rate set screw. It has no
stop and may get out of its housing and provoke a sample leak.
Содержание POLYMETRON 9210
Страница 1: ...Operator Manual POLYMETRON 9210 221 192 010 Revision J 20 11 2007...
Страница 2: ......
Страница 10: ......
Страница 11: ...Page 5 SILKOSTAT 9210 1 Presentation of the instrument Figure 1...
Страница 18: ......
Страница 28: ......
Страница 74: ......
Страница 80: ......
Страница 114: ......
Страница 115: ......
Страница 116: ......