Short
How to disinfect?
Please help us to avoid errors! Copy this sheet and put it up in the cleaning department. Make indications here
on your disinfectant (list of tested disinfectants: www
.haag-streit.com)
Maintain the tonometer
3
5
6
See also: Appendix B.3
Recommendation: Use the
Haag-Streit 'DESINSET'
2.4 Accessories (page 13)
• sensitive feeler arm!
Allowed
• mild soap
• Do not immerse
tonometer in solution!
• No corrosive agents
• No direct spraying
The use of the fol-
lowing products /
procedures for di-
sinfection will de-
grade the tip and
cause it to fail:
• Alcohol
• Acetone
• UV radiation
• Sterilization
• Immersing in
fluid for more
than 1 hour
• Temperature
above 60 °C
Remove
carefully measuring prism from holder
Clean
30 - 60 seconds
running,
cold water
Disinfect
Rinse
min. 0 - max. 60 minutes
running,
cold drinking water
Dry
With a one-way tissue
clean,
soft
Store
Place into container
clean,
dry
•
Clean tonometer only with a dampened cloth
•
Use soap only
Minutes
%
5