5
© Haag-Streit AG, CH-3098 Koeniz, Switzerland
1500.7006000.04101
Sodium Hypochlorite
(Household Bleach)
0 minutes
10% aqueous solution
Haag-Streit USA INC
3535 Kings Mills Road
Mason / Ohio 45040
Phone ++ 1 513 336 7255
Fax
++ 1 513 336 7260
Toll Free 800 787 5426
eMail [email protected]
internet www.haag-streit-usa.com
Short
How to disinfect? in the USA
Maintain the tonometer
3
5
6
See also: Appendix B.3
• sensitive feeler arm!
Allowed
• mild soap
See page 43
for more in-
formation re-
garding the
use of alco-
hol for disin-
fection.
• Do not immerse
tonometer in solution!
• No corrosive agents
• No direct spraying
The use of the fol-
lowing products /
procedures for di-
sinfection will de-
grade the tip and
cause it to fail:
• Alcohol
• Acetone
• UV radiation
• Sterilization
• Immersing in
fluid for more
than 1 hour
• Temperature
above 60 °C
•
Clean tonometer only with a dampened cloth
•
Use soap only
Hydrogen Peroxide
0 minutes
3 % aqueous solution
or
Rinse
rinse thoroughly
running,
cold drinking water.
Note: rinsing time in water may not exceed 60 minutes
Dry
With a one-way tissue,
clean and
soft.
Store
Place into container,
clean and
dry.
Remove
measuring prism carefully from holder.
Clean
Wipe prisms clean before rinsing for...
30 - 60 seconds in
running,
cold water.
Disinfect
5