B.3 Cleaning / disinfecting of measuring prisms
• Tonometer measuring prisms are not dispatched in a disinfec-
ted state and should always be disinfected before use.
• Thorough cleaning under running cold water during 30 to 60
seconds is mandatory before undertaking disinfecting routines.
For more intensive cleaning in case of strong contamination
carefully rub the prism with a cotton-wool ball under running
water. In addition only mild soap may be used.
• For disinfecting purposes we recommend an aqueous 'Pan-
tasept
®
' solution of 3 % during 10 minutes. Refer to the latest
versions of the manufacturer's instructions and sales literature
for details of the exact concentration, effecting time and appli-
cation of disinfectant.
Alternative agents: only those products which are listed on
Haag-Streit AG internet homepage (www.haag-streit.com)
may be used.
• After disinfection the measuring prisms must be rinsed tho-
roughly at least 10 to 60 minutes under running cold drinking
water. Dry them with a clean soft one-way tissue and store
them in a clean dry container.
• Only measuring prisms, free of any residue, should be used.
Residue may cause provoking and caustic reactions on the
patient's eye.
Not allowed!
Measuring prisms can get tension rips
because of:
• Alcohol
• Acetone
• UV radiation
• Sterilization
• Immersing in fluid for more than 1 hour
• Temperature above 60 °C
B. Accessories disinfecting / storing
For efficient disinfecting routines we recommend our DESIN
-
SET (9) which comprises:
• cleaning- and
• storage containers
• 'Pantasept
®
' disinfecting solution