![Haag-Streit Goldmann AT 870 Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/haag-streit/goldmann-at-870/goldmann-at-870_instruction-manual_575787025.webp)
5
© Haag-Streit AG, CH-3098 Koeniz, Switzerland
1500.7006000.04101
Position too much at left / right
The upper semi circular ring appears partly
(17)
The measuring prism is not focused on the eye, the eye is
too far on the right.
Move the slit lamp to the right
with the guide lever.
The upper semi circular ring appears completely - the
lower ring partly
(18)
The measuring prism is not focused on the eye, the eye is still
too far on the right.
Move the slit lamp to the right
with the guide lever.
The lower semi circular ring appears completely - the
upper ring partly
(19)
The measuring prism is not focused on the eye, the eye is still
too far on the left.
Move the slit lamp to the left
with the guide lever.
The lower semi circular ring appears partly
(20)
The measuring prism is not focused on the eye, the eye is
too far on the left.
Move the slit lamp to the left
with the guide lever.
Correct position!
(21)
The edges of both semi circular rings meet exactly in the cen-
tre of the ocular.
17
18
19
20
21