4.2 Setting the weight compensation facility
Turn the control lever (27) to its lowest position and then slacken it slightly by a
quarter turn. Swivel the microscope and illumination to the side. Apply 1-3 springs
depending on the accessory.
4.3 Switching on the compensation facility
Turn anticlockwise until the screws (12) are completely released.
4.4 Switching off the compensation facility
Turn the screws (12) clockwise until you meet resistance. Verify whether the
microscope arm springs back downwards if you push it upwards with your hand.
This will only happen if the load is already at maximum. Generally, as many
counterbalance springs should be deactivated as necessary until this spring action
occurs. The weight compensation facility is set correctly once the illumination and
microscope with the mounted accessories weigh slightly more than the
counterbalance springs.
4.5 Regulating the clearance of the slit width controls
The small screw in the centre of the right control knob (A)
allows you to regulate the friction of the turning movement
of these adjusting knobs. Turning it slightly to the right (in)
makes it harder, turning it left (out) makes it easier. It
should at least be set so hard that the slit cannot close on
its own.
4.6 Attaching/removing a camera
78. Lens barrel
79. Mirror and diaphragm component
80. Camera
81. Eyepiece
Remove the camera cover, attach the spacer ring to the camera and place the
camera on the lens barrel. If the camera is removed, the connection to the lens
barrel should be closed off with the cap intended for this purpose.
NOTE!
Consult your Haag-Streit representative for information on the supported
cameras and their connection rings.
5 Commissioning
Connect the PS-LED power supply and FU 01 flash unit to the mains.
5.1 Switching on the device
• Throw the rocker switch on the PS-LED power supply and the main switch on
the FU 01 flash unit 01 I = ’ON’. The green lamps in the rocker switches light up
when the device is switched on.
• Turn the rotating knob on the illumination control to a position between '1' and
'10'.