3.10 Background illumination
58. Background illumination (swiveling)
3.11 Power supply
NOTE!
• Observe the respective Haag-Streit instructions for use. For further
information, please contact your Haag-Streit representative.
• This device may only be operated with PS-LED and PS-LED HSM
901 power supplies from Haag-Streit.
4 Device assembly / installation
WARNING!
• Do not modify this device without authorization of the manufacturer.
Installation and repairs may only be performed by trained
specialists.
• Contact your Haag-Streit representative for installation, repairs and
modification work on the system. The contact details are available at
www.haag-streit.com.
• Only original Haag-Streit spare parts may be used.
59. Medically approved isolating
transformer
60. Mains
61. Instrument table
62. Camera power supply
63. Printer
64. Screen
65. Personal computer
66. Illumination control IC 01
67. PS-LED or PS-LED HSM 901
68. Flash unit FU 01
69. Optional power input (also
possible with battery operation)
70. Optional camera port (saving to a
card is also possible)
71. Galvanic isolation (EN 60601-1)
72. Headrest
73. Release key
74. LED illumination
75. Mirror and diaphragm component
76. Camera
77. Local network
The initial installation of the device is conducted by an authorized service technician.
The following adjustments can be conducted by the end user:
• Regulating the clearance of the slit width controls
• Attaching/removing a camera
• Adjusting the weight compensation
4.1 Instrument base with weight compensation facility
The weight of additional accessories mounted on the microscope can be
compensated using counterbalance springs. This keeps the height adjustment of the
slit lamp easy.