background image

15

DEUTSCH

ENGLISH 

FRANÇAIS

ITALIANO

ESPAÑOL

NEDERLANDS

PORTUGUÊS

SVENSKA

DEUTSCH

ENGLISH

FRANÇAIS

ITALIANO

ESPAÑOL

NEDERLANDS

PORTUGUÊS

SVENSKA

© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7006000‑04270 – 2022 – 09

Bands are no longer in contact (16)

Pressure is too high.

Lower the pressure by rotating the tonometer knob

in the opposite direction.

Correct setting (17)

The inside borders of the fluorescein bands are con-

tacting each other.

7 Technical data

7.1 Applanation tonometer AT 900

Measuring force

generated

through leverage weight.

Installation

AT 900 model R

Can be fitted on the peg on the microscope.

AT 900 model T

Can be fitted on the guide plate on the swivelling

peg for the microscope and illumination arm.

AT 900 model BQ Assembly on the microscope arm.

Measurement range

0 ― 80 mm Hg

Measurement devi-

ation

The measurement deviation in the measuring prism is in the measur-

ing range from 0 – 58.84 mN and amounts to a maximum of ±1.5%

and to a minimum of ±0.49 mN of the nominal value.

Backlash width

≤0.49 mN

Net weight

AT 900 model R

0.73 kg (without accessories)

AT 900 model T

0.49 kg (without accessories)

AT 900 model BQ 0.78 kg (without accessories)

AT 900 model R

AT 900 model T

AT 900 model BQ

Содержание AT 900

Страница 1: ...GLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7006000 04270 2022 09 INSTRUCTIONS FOR USE Applanation tonometer AT 900 AT 870 27 Edi...

Страница 2: ...ly with the regulations in these instructions for use we can guarantee reliable and trouble free use of our product WARNING Applanation tonometry may only be conducted by appropriately trained and qua...

Страница 3: ...ment 8 2 4 Clinical benefit 8 3 Introduction 9 3 1 Overview 9 4 Device assembly installation 9 4 1 AT 900 model R 9 4 2 AT 900 model T 9 4 3 AT 900 model BQ 9 4 4 Tonometer AT 870 9 4 5 Which slit lam...

Страница 4: ...ing shows traces of im proper handling or damage If this is the case notify the transport company that delivered the goods to you Unpack the device together with a representative of the transport comp...

Страница 5: ...hanges the biomechanical properties of the cornea The biomechanical properties of the cornea have an effect on the measurement readings 1 6 Disinfection NOTE The applanation tonometer can but does not...

Страница 6: ...The operator accepts all liability for the use of other disinfection agents Single use prisms must be replaced after every examination see separate instructions for use The above mentioned instruction...

Страница 7: ...the product This would void all guarantee claims Continued use of a damaged product may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does not gran...

Страница 8: ...2 3 1 Contraindications The list of known contraindications includes but is not limited to the following medic al conditions Active corneal infections Corneal epithelial defects Ocular chemical burn...

Страница 9: ...trained specialists 4 1 AT 900 model R A pivot is required for use of the Applanation Tonometer AT 900 model R on the BM 900 slit lamp It is assembled as follows 1 Unscrew the mounting screw found at...

Страница 10: ...ts from the bottom side into the beam path of the microscope and the illumination The Applanation Tonometer AT 900 model BQ has a working position and two resting positions To set the correct angle of...

Страница 11: ...red IOP A normal corneal thickness is considered to fall within the range of 530 to 560 micrometre If it is suspected that measuring results may be inaccurate please conduct a functional verification...

Страница 12: ...through observation with the wrong eyepiece adjustment can lead to false results 6 6 Measuring correctly NOTE Make sure that the measuring arm of the tonometer can move freely and is not hindered by t...

Страница 13: ...he measuring prism is touching the eye only intermittently If the patient pulls back even further the fluorescein rings will disappear com pletely Instruct the patient to adopt the correct position On...

Страница 14: ...mp up wards Nearly entire circle above incomplete circle below 11 Measuring prism not centred on the eye eye still too high Using the control lever move the slit lamp up wards Two incomplete circles t...

Страница 15: ...ght Installation AT 900 model R Can be fitted on the peg on the microscope AT 900 model T Can be fitted on the guide plate on the swivelling peg for the microscope and illumination arm AT 900 model BQ...

Страница 16: ...evice should be cleaned weekly as described We recommend having the device checked annually by an authorised service tech nician 8 2 Cleaning and disinfection Tonometer housing and sensor arm can if r...

Страница 17: ...AT 870 Check the drum position 0 Check position 0 05 The zero line on the measuring drum is moved downwards opposite the index by a single marking width The feeler arm must move by itself all the way...

Страница 18: ...suring drum s calibration mark 6 down wards opposite the index by a interval The feeler arm must move towards the examiner Check position 6 1 The check point is at 6 1 Move the measuring drum s calibr...

Страница 19: ...on of the patient The feeler arm must move towards the patient Check position 1 95 Move calibration mark 2 on the measuring drum downwards opposite the fixed index by a width of one marking With light...

Страница 20: ...n of the measuring prism and the separate instructions for use 9 3 Tonosafe disposable prisms Tonosafe disposable prisms are a comfortable and ef fective solution to reduce the risk of cross infection...

Страница 21: ...ould you have any further questions please contact your Haag Streit representative at www haag streit com haag streit group contact haag streit distributors distributors HAAG STREIT AG Gartenstadtstra...

Отзывы: