• Check position 1.95
With a drum position of 1.95, the feeler arm must move from the free-movement
area toward the limit stop in the direction of the examiner.
• Check position 2.05
With a drum position of 2.05, the feeler arm must move from the free-movement
area toward the limit stop in the direction of the patient.
Check the drum position 6
Set the weight rod to 6 on the scale, the longer part points to the examiner.
• Check position 5.9
The check point is at 5.9. Move the measuring drum's calibration mark 6 down-
wards, opposite the index, by a ½ interval. The feeler arm must move towards
the examiner.
• Check position 6.1
The check point is at 6.1. Move the measuring drum's calibration mark 6 up-
wards, opposite the index, by a ½ interval. The feeler arm must move towards
the patient.
8.4 Checking the Tonometer AT 870
NOTE!
This check must be carried out once a month. If the check yields erro-
neous results, please check the following points:
1. Is the measuring prism properly positioned?
2. Is the control weight correctly set?
3. Repeat the check.
Defective devices must be sent immediately to your Haag-Streit representative.
NOTE!
Haag-Streit offers a repair and service package. For further information,
please contact your Haag-Streit representative directly.