© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7006000‑04270 – 2022 – 09
1 Safety
DANGER!
Failure to comply with these instructions may result in material damage
or pose a danger to patients or users.
WARNING!
These warnings must absolutely be complied with to guarantee safe op-
eration of the product and to avoid any danger to users and to patients.
NOTE!
Important information, please read carefully.
1.1 Comments on these instructions for use
NOTE!
In these instructions for use the point is used as decimal separator.
1.2 Ambient conditions
Transport
Temperature
−40 °C ... +70 °C
Air pressure
500 hPa ... 1060 hPa
Relative humidity
10 % ... 95 %
Storage
Temperature
−10 °C ... +55 °C
Air pressure
700 hPa ... 1060 hPa
Relative humidity
10 % ... 95 %
Use
Temperature
+10 °C ... +35 °C
Air pressure
800 hPa ... 1060 hPa
Relative humidity
30 % ... 90 %
1.3 Shipment and unpacking
• Before unpacking the device, check whether the packaging shows traces of im-
proper handling or damage. If this is the case, notify the transport company that
delivered the goods to you.
• Unpack the device together with a representative of the transport company.
Make a report of any damaged parts. This report must be signed by you and by
the representative of the transport company.
• Leave the device in the packaging for a few hours before unpacking it (condens-
ation).
• Check the device for damage after it is unpacked.
• Return defective devices in the appropriate packaging.
• Store packaging material carefully so that it can be used for potential returns or
when moving.
NOTE!
Check the calibration of the device before first use according to the in-
structions in the 'Maintenance' chapter.
1.4 Installation warnings
WARNING!
Do not modify this device without authorization of the manufacturer. In-
stallation and repairs may only be performed by trained specialists.
NOTE!
• When performing assembly on devices from other manufacturers,
the connecting dimensions of the diverse tonometer models are to
be taken into account.
• Check: Are the connection parts properly positioned (tonometer on
the slit lamp, measuring prisms)?