background image

12

DEUTSCH

ENGLISH 

FRANÇAIS

ITALIANO

ESPAÑOL

NEDERLANDS

PORTUGUÊS

SVENSKA

© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7006000‑04270 – 2022 – 09

AT 900 model R and AT 900 model BQ

1. Swing the illumination apparatus to the left.

2. Release the tonometer from the dwell position to the right of the microscope,

and swing it forward until it locks in the measuring position.

3. From the left, bring the illumination apparatus into contact with the tonometer

bearer arm. This is the only illumination position in which both the patient’s left

and right eyes can be easily examined (no 60° position). This arrangement fa-

cilitates the splaying of the patient’s eyelids, should this be necessary for

measurement. The illumination of the applanated surface through the measur-

ing prism is practically reflection-free.

Observation:

with AT 900 model R in the left eyepiece

with AT 900 model BQ in the right eyepiece

AT 900 model T

For an examination through the tonometer’s left or right eyepiece, the angle

between the illumination apparatus and the microscope should be approx. 60° so

that the image is bright and reflection-free. Alternatively: lighting through the meas-

uring prism at approx. 10°.

AT 870

1. Swing the sensor arm with the measuring prism into the beam path of the mi-

croscope and the illumination.

2. The angle between the illumination apparatus and the microscope should be

approximately 60° so that the image appears bright and free of reflections. Al-

ternatively: lighting through the measuring prism at approx. 10°.

3. Set the measuring drum to the first calibration line.

4. Set a mid-level illumination intensity.

WARNING!

Measurements through observation with the wrong eyepiece adjustment

can lead to false results.

6.6 Measuring correctly

NOTE!

Make sure that the measuring arm of the tonometer can move freely and

is not hindered by the patient's eyelashes, beard, etc.

1. Immediately before taking measurements, the patient should close his eyes

briefly so that the cornea becomes sufficiently moistened with fluorescein-im-

pregnated tear fluid.

2. By moving the slit lamp, the measuring prism comes into contact with the

centre of the cornea over the pupillary area.

3. During contact, the cornea's limbus takes on a bluish glow. This glow can be

best observed with the naked eye from the opposite side of the illumination ap-

paratus.

4. When the limbus glows, stop moving the slit lamp immediately.

5. After contact is made, viewing is conducted through the microscope. The uni-

form pulsation of the two semi circular fluorescein bands, which could be of dif-

ferent sizes in drum setting 1 depending upon inter-ocular pressure, shows

that the tonometer is in the right measuring position.

6. Any necessary corrections are done using the slit lamp control lever, until the

flattened surface is observed in the form of two semicircles of similar size in

the middle of the visual field (A).

7. Smaller changes in the depth of the slit lamp using the control lever do not af-

fect the size of the semicircles.

8. The pressure on the eye is increased by turning the tonometer knob until the

inner borders of both fluorescein bands just touch = correct setting (B).

9. When the eye pulsates, both semi circles cross over each other.

10. The width of the fluorescein band around the contact point of the measuring

prism should be about 1/10 of the diameter of the applanation surface (0.3

mm).

11. Reading the scale value:

• Reading the value

• and multiplying by 10

• yields the inter-ocular pressure in mm Hg

Содержание AT 900

Страница 1: ...GLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7006000 04270 2022 09 INSTRUCTIONS FOR USE Applanation tonometer AT 900 AT 870 27 Edi...

Страница 2: ...ly with the regulations in these instructions for use we can guarantee reliable and trouble free use of our product WARNING Applanation tonometry may only be conducted by appropriately trained and qua...

Страница 3: ...ment 8 2 4 Clinical benefit 8 3 Introduction 9 3 1 Overview 9 4 Device assembly installation 9 4 1 AT 900 model R 9 4 2 AT 900 model T 9 4 3 AT 900 model BQ 9 4 4 Tonometer AT 870 9 4 5 Which slit lam...

Страница 4: ...ing shows traces of im proper handling or damage If this is the case notify the transport company that delivered the goods to you Unpack the device together with a representative of the transport comp...

Страница 5: ...hanges the biomechanical properties of the cornea The biomechanical properties of the cornea have an effect on the measurement readings 1 6 Disinfection NOTE The applanation tonometer can but does not...

Страница 6: ...The operator accepts all liability for the use of other disinfection agents Single use prisms must be replaced after every examination see separate instructions for use The above mentioned instruction...

Страница 7: ...the product This would void all guarantee claims Continued use of a damaged product may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does not gran...

Страница 8: ...2 3 1 Contraindications The list of known contraindications includes but is not limited to the following medic al conditions Active corneal infections Corneal epithelial defects Ocular chemical burn...

Страница 9: ...trained specialists 4 1 AT 900 model R A pivot is required for use of the Applanation Tonometer AT 900 model R on the BM 900 slit lamp It is assembled as follows 1 Unscrew the mounting screw found at...

Страница 10: ...ts from the bottom side into the beam path of the microscope and the illumination The Applanation Tonometer AT 900 model BQ has a working position and two resting positions To set the correct angle of...

Страница 11: ...red IOP A normal corneal thickness is considered to fall within the range of 530 to 560 micrometre If it is suspected that measuring results may be inaccurate please conduct a functional verification...

Страница 12: ...through observation with the wrong eyepiece adjustment can lead to false results 6 6 Measuring correctly NOTE Make sure that the measuring arm of the tonometer can move freely and is not hindered by t...

Страница 13: ...he measuring prism is touching the eye only intermittently If the patient pulls back even further the fluorescein rings will disappear com pletely Instruct the patient to adopt the correct position On...

Страница 14: ...mp up wards Nearly entire circle above incomplete circle below 11 Measuring prism not centred on the eye eye still too high Using the control lever move the slit lamp up wards Two incomplete circles t...

Страница 15: ...ght Installation AT 900 model R Can be fitted on the peg on the microscope AT 900 model T Can be fitted on the guide plate on the swivelling peg for the microscope and illumination arm AT 900 model BQ...

Страница 16: ...evice should be cleaned weekly as described We recommend having the device checked annually by an authorised service tech nician 8 2 Cleaning and disinfection Tonometer housing and sensor arm can if r...

Страница 17: ...AT 870 Check the drum position 0 Check position 0 05 The zero line on the measuring drum is moved downwards opposite the index by a single marking width The feeler arm must move by itself all the way...

Страница 18: ...suring drum s calibration mark 6 down wards opposite the index by a interval The feeler arm must move towards the examiner Check position 6 1 The check point is at 6 1 Move the measuring drum s calibr...

Страница 19: ...on of the patient The feeler arm must move towards the patient Check position 1 95 Move calibration mark 2 on the measuring drum downwards opposite the fixed index by a width of one marking With light...

Страница 20: ...n of the measuring prism and the separate instructions for use 9 3 Tonosafe disposable prisms Tonosafe disposable prisms are a comfortable and ef fective solution to reduce the risk of cross infection...

Страница 21: ...ould you have any further questions please contact your Haag Streit representative at www haag streit com haag streit group contact haag streit distributors distributors HAAG STREIT AG Gartenstadtstra...

Отзывы: