optional settings: administrator options
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh [email protected] created 30.01.2015 printed 30.01.2015 page: 156 of 216
OP93
Total time of operation (h)
= standby time including time of use
of the motor / running belt
Note:
It is not allowed to change the total
time of operation or the total
distance to a value which does not
match the real value of the device.
§
and
$
report: hours of operation
%
indicates h
Difference to user option 03: the displayed value can also be
changed.
OP94
Total time of use (h)
= time of use of motor / running belt
only
Note:
It is not allowed to change the total
time of operation or the total
distance of the running machine to a
value which does not match the real
value of the device
§
and
$
report: hours of operation
%
reports: h
Difference to user option 04: the displayed value can also be
changed.
OP95
MCU serial number
Report and entering of 8-digit serial number with the following format:
YYMM NNNN
YY:
2-digit number of year of production lot
MM:
month of production lot
NNNN: 4-digit serial number
OP98
Setting of limitation of usage time
Secret option
OP99
Test profile 99 for endurance testing
in the factory / service
With this option a special test profile (profile number 99) for
endurance testing in the factory or service can be activated. This test
profile is not for normal use and must be deactivated again after
using. If the device is switched off and on again, profile number 99 is
deactivated automatically.
Details of test profile 99:
Elevation from 0 to maximum elevation every 20 min.
Speed 3.0 km/h for 2 hours
(for ladder ergometer: 3.0 m/min for 2 hours)
Speed 5.0 km/h for 3 hours
(for ladder ergometer: 5.0 m/min for 3 hours)
Speed 8.0 km/h for 8 hours
(for ladder ergometer: 8.0 m/min for 8 hours)
Repeat after 13 hours.
Adjustable settings:
0 = test profile 99 locked / de-activated
1 = test profile 99 activated