PCS-1000 User Manual
72
[SENSe:]VOLTage:AC:AVERage:COUNt
Set
Query
Description
This query will set or return the average count
setting for AC current.
Syntax
Query Syntax
[SENSe:]VOLTage:AC:AVERage:COUNt <NR1>
[SENSe:]VOLTage:AC:AVERage:COUNt?
Return Parameter <NR1>
The average count setting for AC voltage.
1~10, 20, 30, 40, 50, 60, 70, 80, 90, 100
Query Example
VOLT:AC:AVER:COUN?
>10