COMMUNICATION INTERFACE
71
Parameter /
Return Parameter
<Range>
Sets the voltage range in volts. The unit
will automatically choose the closest
range that is programmed.
DC Range <NRf>: 0.0000001 ~ 1050
AC Range <NRf>: 0.0000001 ~ 600
AUTO
Sets the range to AUTO.
Example
VOLT:RANG AUTO
Sets the voltage range to auto.
Note
The range that is returned is the base voltage unit. See
the table below:
Unit
Voltage Range
1000
1000VDC
600
600ACV
100
200V
10
20V
1
2V
0.1
200mV
[SENSe:]VOLTage:DC:AVERage:COUNt
Set
Query
Description
This command will set or return the average count
setting for DC voltage.
Syntax
Query Syntax
[SENSe:]VOLTage:DC:AVERage:COUNt <NR1>
[SENSe:]VOLTage:DC:AVERage:COUNt?
Parameter /
Return Parameter
<NR1>
The average count setting for DC voltage.
1~10, 20, 30, 40, 50, 60, 70, 80, 90, 100
Query Example
VOLT:DC:AVER:COUN?
>10
The average count setting for DC voltage is 10.