COMMUNICATION INTERFACE
63
Example
CONF:CURR 20
Sets the current mode to DC and the range to 30A
Example
CONF:CURR
Sets the current mode to DC. The range is not
changed.
CONFigure:CURRent:AC
Set
Description
This command will set current mode to AC and set
the range. If the range is not specified, then it will
not change.
Syntax
CONFigure:CURRent:AC [<Range> | AUTO]
Parameter
<Range>
Current range <NRf>: 0.00000001~305
Current range. The unit will
automatically be set to the closest range.
AUTO
Autorange; Only applicable for the ≤3A
ranges.
Autorange is not supported for the 30A
and 300A ranges.
Example
CONF:CURR:AC 100
Sets the current mode to AC and the range to 300A.
Example
CONF:CURR:AC
Sets the current mode to AC. The range is not
changed.
CONFigure:VOLTage
Query
Description
The CONFigure:VOLTage query will return the
voltage mode and the voltage range unit.
Query Syntax
CONFigure:VOLTage?
Return Parameter <string>
Returns the voltage mode and range unit.