OPERATION
37
Setting the AD Speed
Background
The ADC IC speed has a number of settings.
The higher the setting, the lower the accuracy
and resolution of the meter.
Range:
Seconds (resolution):
7 (6½ digits), 30 (5½ digits),
100 (4½ digits)
Steps
1.
Long push the
Func
key.
The function menu will appear.
2.
Use the
◄
Func
►
keys to navigate to the
AD
SPEED
menu.
3.
Use the
▼
Select
▲
keys to select the AD speed.
Press the
Enter
key to set the speed.
By default the AD Speed is set to 7 (6½ digits).
4.
Use the
◄
Func
►
keys to navigate to the
SAVE FUNC SET
menu item. Press the
Enter
key to save.
Note
To exit without saving, navigate to the
EXIT FUNC
SET menu using the
◄
Func
►
keys and press the
Enter key to exit without saving any settings.