APPENDIX
151
Resistance (2-wire resistance and 4-wire resistance)
Measurement method: 2-wire resistance or 4-wire resistance.
Open-circuit voltage: Approximately 7.5 VDC.
Input protection: 500Vpeak on all ranges.
Diode
Measurement method: 1mA ±2% constant current source.
Open-circuit voltage: Approximately 7.5 VDC.
Input protection: Input protection of 500V peak.
Rate
Additional Rate Error Count
Med
50
Fast
200
Continuity
Measurement method: 1mA ±2% constant current source.
Open-circuit voltage: Approximately 7.5 VDC.
Input protection: Input protection of 500V peak.
Continuity threshold: 0
Ω
~ 2000
Ω.
Threshold step: 1
Ω.
Rate
Additional Rate Error Count
Med
60
Fast
200