GDM-8352 User Manual
120
[SENSe:]FREQuency:INPutjack
Assigns an input terminal for the frequency function.
Parameter: (0|1|2) 0=volt, 1=1A, 2=10A
Example: SENS:FREQ:INP 0
Sets the input jack to the Volt input terminal.
[SENSe:]FREQuency:INPutjack?
Returns the assigned input terminal used for the frequency
function.
Return Parameter: VOLT, 1A, 10A
[SENSe:]PERiod:INPutjack
Assigns an input terminal for the period function.
Parameter: (0|1|2) 0=volt, 1=1A, 2=10A
Example: SENS:PER:INP 0
Sets the input jack to the Volt input terminal.
[SENSe:]PERiod:INPutjack?
Returns the assigned input terminal used for the period function.
Return Parameter: VOLT, 1A, 10A
[SENSe:]CONTinuity:THReshold
Sets the continuity threshold in ohms.
Parameter: <NRf> (0 ~ 2000)
Example: SENS:CONT:THR 500
Sets the continuity threshold to 500 ohms.
[SENSe:]CONTinuity:THReshold?
Returns the continuity threshold.
Return Parameter: <NR1> (0~2000)