GDM-8352 User Manual
114
MEASure:CURRent:AC?
Returns the AC current measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:CURR:AC?
> +0.00872E-02
Returns the AC current measurement as 0.0872mA.
MEASure:CURRent:DCAC?
Returns the DC+AC current measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:CURR:DCAC?
>+0.01245E-02
Returns the DC+AC current measurement as 0.1245 mA.
MEASure:RESistance?
Returns the 2W resistance measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:RES?
> +1.03
Returns the 2W measurement as
1.00156kΩ.
MEASure:FRESistance?
Returns the 4W resistance measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:FRES?
> +1.03
Returns the 4W measurement as
1.11365kΩ.
MEASure:FREQuency?
Returns the frequency measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:FREQ?
> +1.03
Returns the frequency (1.00123kHz).