APPENDIX
149
AC Voltage (AC Coupling Mode/AC + DC Coupling Mode)
Measurement method: AC coupled true RMS - measure the AC
component with up to 400 VDC bias on any range.
Crest Factor: Maximum 3 at full scale.
Input Impedance: 1 MΩ ± 2% in parallel with <100 pF on all ranges.
Maximum input voltage: 750 Vrms on all ranges.
Input protection: 1200V peak on all ranges with gas discharge.
Rate
[1] Frequency
Med
>20Hz
Fast
>200Hz
Rate
Range
Accuracy
20 Hz to
45 Hz
45 Hz to
10 kHz
10 kHz to
30 kHz
30 kHz to
100 kHz
Med
200.000mV 1% + 200
0.3% + 400
1.5% +800
5% + 1200
2.00000 V 1% + 200
0.2% + 400
1% +400
3% + 800
20.0000 V 1% + 200
0.2% + 400
1% +400
3% + 800
200.000 V 1% + 200
0.2% + 400
1% +400
3% + 800
750.00 V 1% + 200
0.2% + 400
1% +400
3% + 800
Fast
200.000mV
-
0.3% + 1000 1.5% +1000
5% + 1500
2.00000 V
-
0.2% + 500
1% +500
3% + 1000
20.0000 V
-
0.2% + 500
1% +500
3% + 1000
200.000 V
-
0.2% + 500
1% +500
3% + 1000
750.00 V
-
0.2% + 500
1% +500
3% + 1000
*The accuracy of ACV+DCV is equal to ACV’s with 10 more digits added.
[1] The accuracy of the AC voltage measurement is guaranteed only when the
signal being measured has frequencies higher than what is listed here.