APPENDIX
145
Resistance
[1]
[2]
Range
Resolution
full scale
Current
source
Accuracy (4W)
200.000
Ω
0.001
Ω
239.999
1mA
0.05% + 8
2.00000k
Ω
0.00001k
Ω
2.39999
1mA
0.05% + 5
20.0000k
Ω
0.0001k
Ω
23.9999
100
μ
A
0.05% + 5
200.000k
Ω
0.001k
Ω
239.999
10
μ
A
0.05% + 5
2.00000M
Ω
0.00001M
Ω
2.39999
1
μ
A
0.05% + 5
10.0000M
Ω
0.0001M
Ω
11.9999
0.5
μ
A
0.3% + 5
100.000M
Ω
0.001M
Ω
119.999
0.5
μ
A//10M
Ω
3.0% + 8
[1] Specifications are for 4-wire resistance measurement, or 2-wire resistance
measurement using the REL function. Without the REL function, add 0.2
Ω
additional error when using 2-wire resistance measurement.
[2] When measuring resistances greater than 500k
Ω
, please use shielded test
leads to eliminate the noise interference that may be induced by standard test
leads.
AC Voltage
[1] [2]
Range Resolution
Full
Scale
20 Hz to
45 Hz
45 Hz to
10kHz
10 kHz to
30 kHz
30 kHz to
100 kHz
200.000mV 0.001mV
239.999
1% + 100
0.3% + 100
1.5% +300 5% + 300
2.00000 V 0.00001V
2.39999
1% + 100
0.2% + 100
1% +100
3% + 200
20.0000 V
0.0001V
23.9999
1% + 100
0.2% + 100
1% +100
3% + 200
200.000 V
0.001V
239.999
1% + 100
0.2% + 100
1% +100
3% + 200
750.00 V
[3]
0.01V
765.00
1% + 100
0.2% + 100
1% +100
3% + 200
[1] Specifications are for sine wave inputs that are greater than 5% range.
[2] Rate in Fast, Input ACV Frequency > 200Hz.
[3] The specifications are guaranteed to an input of 750V. A beeping alarm will