User Manual
the data on the image will be highlighted by a black background so that the data is
easy to be recognized.
•
Laser:
To turn on or off the laser locator when viewing live image
•
Audio Alarm:
To turn on or off the audio alarm for temperature exceeding the
preset value
•
Voice Annotat:
To enable or disable voice annotation when saving live images
•
Temp Unit
: To switch between Celsius and Fahrenheit
•
Mean
: To define the frozen image as mean of different frames.
•
Press Button “Menu” or “OK” to confirm the parameter setting and exit “Setup
Analysis” menu.
•
Press Button “C’ to cancel the parameter setting and exit “Setup Analysis” menu.
Press “C” again to exit main menu.
Sub Menu Target
•
Press Button “Menu” or “
>
” to enter the option “Setup”. Press Button “ ” to
∨
highlight “Target”
•
Press Button “Menu” or “OK” to enter the “Target Setup” option
•
The setup parameters menu will then appear. Press “ ”and“ ”
∧
∨
to move between
each setting, press Button “
<
”and “
>
” to alter the values
•
Filter:
Set it as 1 or 2 according to its corresponding temperature range.
•
Emissivity
: Set emissivity value between 0.01 and 0.99
•
Relative Humidity
: Set humidity percentage value between 1 and 99
•
Distance
: Set distance from target (meters)
•
Alarm
: Set scene temperature alarm threshold for audible alarm