Geophysical Survey Systems, Inc.
SIR® 30
Manual
MN 93-101 Rev E
39
1
To calibrate the survey wheel, lay out a measuring tape and measure out a calibration distance.
2
Enter the distance in the DISTANCE dialog box. Position the midpoint of the antenna at zero and click
START.
3
Move the antenna to the end of your measured distance and then click STOP. You will see the Tick
Count change while the survey wheel is moving.
•
After selecting stop, the User will see a new calibration value displayed. This represents number of
ticks per unit distance the survey wheel encoder produces.
•
The User may repeat this process several times to make sure the calibration value is consistent.
•
When satisfied with the calibration Value click ACCEPT. The User should record this number for
future use.
•
The number of ticks per revolution should not change significantly for GSSI survey wheels
connected to GSSI carts or antennas.
•
If your Survey Wheel\DMI is connected to the axle of a vehicle (GSSI part # 630-B Vehicle
Mounted Survey Wheel encoder) the User should also note the tire pressure. Changes in tire
pressure will lead to small changes in the radius of the tire and a variation in the calibration value.
Note:
All GSSI supplied carts, survey wheels and DMI’s operate in Quadrature mode. A Quadrature
device will count up and down while a pulse device will only count up. If you are using a non-GSSI
supplied device, refer to the manufacturer’s documentation or contact your device’s manufacturer to
find out what type of DMI you have.
RADAR CONFIG > MAINTENANCE > CONFIGURE GPS
This menu allows the User to control the SIR 30 serial COM port settings to communicate with an external
GPS.
1
Select CONFIGURE GPS. This will open the GPS PORT CONFIG menu.
2
The User must configure the SIR 30 input COM port parameters to match the GPS output COM port
parameters. If the GPS output parameters do not match the SIR 30 input parameters, communications
between the SIR 30 and the GPS system will not take place.
3
When the User is satisfied with the GPS port configuration click ACCEPT.
Note:
The SIR 30 accepts a NEMA 0183 GGA string.
Содержание SIR 30
Страница 1: ......
Страница 4: ...Geophysical Survey Systems Inc SIR 30 Manual ...
Страница 6: ......
Страница 90: ...Geophysical Survey Systems Inc SIR 30 Manual MN 93 101 Rev E 84 ...
Страница 102: ...Geophysical Survey Systems Inc SIR 30 Manual MN 93 101 Rev E 96 ...
Страница 120: ...Geophysical Survey Systems Inc SIR 30 Manual MN 93 101 Rev E 114 ...
Страница 122: ...Geophysical Survey Systems Inc SIR 30 Manual MN 93 101 Rev E 116 ...
Страница 144: ...Geophysical Survey Systems Inc SIR 30 Manual MN 93 101 Rev E 138 Figure 168 SIR 30 Cable Pin out ...
Страница 145: ...Geophysical Survey Systems Inc SIR 30 Manual MN 93 101 Rev E 139 Figure 169 Cable Selection Guide ...
Страница 146: ...Geophysical Survey Systems Inc SIR 30 Manual MN 93 101 Rev E 140 Figure 170 Survey Wheel and Model 10 Marker Pin Out ...
Страница 147: ...Geophysical Survey Systems Inc SIR 30 Manual MN 93 101 Rev E 141 Figure 171 Serial Port GPS Pin out ...
Страница 148: ...Geophysical Survey Systems Inc SIR 30 Manual MN 93 101 Rev E 142 ...