ZERO IN NITROGEN
Place the sensor in nitrogen gas and allow time for the sensor to stabilise and the gas to be fully
diffused into the sensor. Click on
‘Nitrogen’
. The sensor will be zeroed assuming a 0ppm CO
2
environment.
ZERO IN AMBIENT CONDITIONS
If there is no calibration gas or nitrogen available, the sensor zero-point can be set in fresh air.
Ambient CO
2
concentrations in fresh air are typically 400ppm. Place the sensor in a fresh air
environment and allow time for the sensor temperature to stabilise, and for the fresh air to be fully
diffused into the sensor.
If there is no value in the ‘Set to’ dialogue box,
when
‘Ambient’
is clicked,
the sensor will be zeroed assuming the ambient environment is 400ppm.
The CO
2
concentration fresh air zero level is programmable over a range from 0ppm to the full scale
of the sensor. The software default is 400ppm as the fresh air CO
2
concentration value. However,
the user can write a different fresh air value to the sensor if desired.
To write a different fresh air CO
2
concentration value to the sensor, set the CO
2
concentration value
to the appropriate value by writing the value in the ‘set to’ box. Click on ‘Am
bient.
ZERO IN A KNOWN GAS CONCENTRATION
Place the sensor in a known gas concentration, and allow time for the sensor temperature to
stabilise, and for the gas to be fully diffused into the sensor. Set the CO
2
concentration value to the
appropriate value by writing the value in the ‘
set to
’
box. Click on
‘Known Gas’.
In all cases the sensor polling mode needs to be switched off during the change. The software will
warn the user if this is not selected. After the change has been made polling mode can be re-
enabled.