![GSS CozIR-Blink Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/gss/cozir-blink/cozir-blink_user-manual_2275520019.webp)
USER GUIDE
Gas Sensing Solutions Ltd.
Revision 1.2, 29 March 2021
Page | 19 For regular updates, sign up at https://gassensing.co.uk
Copyright © 2021 Gas Sensing Solutions Ltd.
CURRENT MEASUREMENT
To
measure current, the sensor must be in ‘Manual Measurement’ mode.
The current measurement function generates a real time graphical display of the current
consumption of the sensor.
To correctly measure average current, the period between power cycles must be specified. This is
done by entering a value in the ‘Measure every seconds’ box. In the example shown, the period is
300 seconds between power on events.
To measure average current, click on
‘Update’
Current (uA) in the Graph Data section.
The ‘Average Current’ displayed in the
lower centre of the screen is the current consumed by the
sensor over a complete 300s power cycle period. Thus, the average current is defined as the current
consumed by the sensor during the measurement period when it is powered up (period defined by
the nPulse setting) divided by the total time between power on events.
Note that the sensor is powered down as soon as the CO
2
reading has been accessed by the host
controller.