USER GUIDE
Gas Sensing Solutions Ltd.
Revision 1.2, 29 March 2021
Page | 10 For regular updates, sign up at https://gassensing.co.uk
Copyright © 2021 Gas Sensing Solutions Ltd.
Method 3 - Variable External Power Supply
The user can also drive the sensor directly with an external power supply, bypassing the on-board
low noise LDO and FET switch. This allows the user to evaluate the behaviour of the sensor with
different supply voltages or power sources. To enable this mode of operation, the following settings
need to be configured.
Connector/Jumper
Setting
Sensor Connection
J5, pin 1
Connect to external GND
GND
J5, pin 2
Connect to external VDD
VDD
J5, pin 4
Connect to external VDD
VDD1
PL1
Remove jumper. Isolates sensor from
rest of the evaluation board
-
PL2
Remove jumper. Isolates sensor from
rest of the evaluation board
-
The evaluation board has pull-up resistors for the I
2
C serial data input/output lines in each of these
modes. They pull-up to either the on-board 3.3V supply, or the external VDD on J5, pin 2 and pin 4.
All the sensor connections are enumerated on connectors, J3, J4 and J5, see figure 1. For a full
description of the sensor connections and functions, please refer to the latest version of the data
sheet at, https://www.gassensing.co.uk/products/.
Note that for the CozIR®-Blink-N to operate correctly in this mode, the external supply connections
on J5, pin 2 and pin 4 must be power-cycled. This means that all power to the sensor must be
removed, and then re-applied. The CozIR®-Blink-N will only take a new reading after it has been
power cycled. The time between a power down and re-application of power should be greater than 4
seconds to ensure the sensor is fully switched off. This will ensure the sensor is properly power
cycled. Failure to allow the sensor adequate time to power down may result in erratic performance.