SLT 5.0 Level and Flow Monitor
POWER INPUT OPTION
9-32VDC
SLT 5.0 Level & Flow Monitors may be ordered factory-configured for 9-32VDC power input.
QUICK BENCH TEST:
Connect Sensor as shown below, then Power. When properly connected figures will show on the large
LCD display. Test operation of the SLT 5.0 by holding the sensor steadily and aiming at a flat, stable
target 12 to 28" (305 to 711 mm) away from the end of the sensor. Allow a few seconds for the SLT 5.0
to lock onto the target before displaying its distance. The SLT 5.0 will now display Range in ft or cm
(factory calibration).
CONNECTIONS:
POWER INPUT: Connect 9-32VDC/0.5 Amps to the + and - terminals. The Power Input GND must be
connected to the nearest Ground pole. A 1 amp fuse in line is recommended.