Page 17
SLT 5.0 Level and Flow Monitor
SIMULATION
Exercises the 4-20mA (0-5V) output, digital display and control relays.
Test
Select
Maximum
and press
to simulate maximum Flow, Range or
Level and to output 20mA (5V) to the analog channel.
Select
Minimum
and press
to simulate minimum Flow, Range or
Level and to output 4mA (0V) to the analog channel.
To simulate and intermediate Flow, Range or Level set
Test
to
Actual
and then enter a value for the
Flow
,
Range
or
Level
. The analog
output and control relays will respond to the simulated value.
Содержание SLT 5.0