Page 11
SLT 5.0 Level and Flow Monitor
CALIBRATION
Press the
to
Calibration
and
to enter. Use
or
to position
before
each menu item and
to enter. When settings are completed press
to store and
return to the Calibration menu.
Min Range
Distance from the sensor face to highest expected level.
Max Range
Distance from sensor to Zero level
MaxVol
For Volume and HRT mode only. Enter the maximum volume of
the vertical or horizontal tank.
Damping
Minimum damping allows fast response to level changes.
Increasing damping slows the SLT 5.0's response to level changes
and is ideal to smooth the display and outputs in turbulent
conditions. Damping value is shown in percent (0-99%). Some
experimentation may be required to select the optimum damping
value. A value of 1% is recommended for most applications. For
fast level changes (up to ½ inch/sec - 13 mm/sec), a Damping
value of 1% is recommended.
LOE Time
Press
and
or
to change the number of seconds without
receiving an echo before the SLT 5.0 displays ECHO LOSS, and
Control relays change state as calibrated under Relay Parameters.
Factory default is 30 seconds and is recommended for most
applications, Minimum is 1 second and maximum is 99 seconds.
Press
from the
Units/Mode
display to return to Menu Selections.
Содержание SLT 5.0
Страница 2: ...Note This page has been left blank intentionally ...
Страница 5: ...Page 5 SLT 5 0 Level and Flow Monitor CONNECTIONS ...
Страница 7: ...Page 7 SLT 5 0 Level and Flow Monitor CALIBRATION MENU ...
Страница 14: ...Page 14 SLT 5 0 Level and Flow Monitor CHANNEL SETUP ...
Страница 20: ...Page 20 SLT 5 0 Level and Flow Monitor SENSOR MOUNTING METHODS ...