H84.0.15.6C-01
Operating manual ResOx 5695
Page 9 of 22
_____________________________________________________
_____________________________________________________________________________
ADR
0 1
,
1 1
,
2 1
,
…
9 1
Base address for serial interface communication
(ex works
setting 01)
DA.,N
[ONC
Analog output is corresponding to concentration in %
P. 02
Analog output is corresponding to partial pressure in hPa or
mmHg
DA[.0
0.0...100.0
%O
2
Measuring value that should correspond to output 0 V
e.g. for,
(ex works setting 0.0 % Vol O
2
)
DA[. 1
0.0...100.0
%O
2
Measuring value that should correspond to output 1 V
e.g. for,
(ex works setting 100.0 % Vol O
2
)
SET
(ORR
Set Corr: Input adjustment
**
11
0FFS
°C bzw. °F
-5.0
°C
... 5.0
°C
or
-9.0
°F .
.. 9.0
°F
The zero point of the temperature measuring is shifted for the
entered value. This can be used to compensate sensor and
instrument deviations
**
OFF
No zero adjustment for temperature measurement (=0.0°)
S[AL
°C bzw. °F
-5.00 ... 5.00
%
The slope of the temperature measurement is corrected by this
value.
This can be used to compensate sensor and instrument
deviations
**
OFF
No slope adjustment for temperature measurement (=0.00)
0FFS
hPa
-20 ... 20
hPa
The zero point of the pressure measuring is shifted for the
entered value.
This can be used to compensate sensor deviations
**
OFF
No zero adjustment for pressure measurement (=0 hPa)
SET
AL
Set Alarm: Configuration of the alarm settings
13
AL. 1
ON
/
NO.SO
Monitoring oxygen: Alarm on with buzzer / Alarm on without
buzzer
OFF
No alarm monitoring for oxygen
AL.,N
[ONC
Monitoring oxygen: Concentration in % Vol O
2
P. 02
Monitoring oxygen: Partial pressure in hPa or mmHg
A 1.LO
e.g.
0.0...100.0
%
Min alarm limit oxygen (not if AL. 1. oFF)
A 1.H,
e.g.
0.0...100.0
%
Max alarm limit oxygen (not if AL. 1. oFF)
AL. 2
ON
/
NO.SO
Temperature monitoring : Alarm on with buzzer / Alarm on
without buzzer
OFF
No alarm monitoring for temperature
A2.LO
-5.0 ...
+
50.0
°C
Min alarm limit temperature (not if AL. 2. oFF)
A2.H,
-5.0 ...
+
50.0
°C
Max alarm limit temperature (not if AL. 2. oFF)
SET
LO66
Set Logger: Configuration of the logger functions
10.3
FVNC
(Y(L
Cyclic
: Automatic logger function
*
STOR
Store
: manual recording
OFF
No logger activated
(Y(L
0:0 1 ... 60:00
Cycle time in [minutes: seconds] for cyclic logger
**
SET
(L0(
Set Clock: Configuration of the real time
14
(L0(
HH:MM
Clock
: set time
hours: minutes
YEAR
YYYY
Year
: set year
DATE
TT.MM
Date
: set date
day: month
(*)
If logger memory contains data sets parameters marked with (*) cannot be called. You
have to clear memory to change these parameters!
(**)
If logger is running parameters marked with (**) cannot be called