H84.0.15.6C-01 

Operating manual ResOx 5695 

Page 9 of 22 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

 

ADR

 

0 1

,

 1 1

,

2 1

,

 

 9 1

 

Base address for serial interface communication 

(ex works 

setting 01)

 

 

 

DA.,N 

[ONC 

Analog output is corresponding to concentration in % 

 

 

P. 02 

Analog output is corresponding to partial pressure in hPa or 
mmHg 

DA[.0 

0.0...100.0

 

%O

2

 

Measuring value that should correspond to output 0 V  
e.g. for, 

(ex works setting 0.0 % Vol O

2

 

 

DA[. 1 

0.0...100.0

 %O

2

 

Measuring value that should correspond to output 1 V  
e.g. for, 

(ex works setting 100.0 % Vol O

2

 

 

SET 

(ORR

 

Set Corr: Input adjustment 

** 

11 

0FFS 

°C bzw. °F

 

-5.0

 °C

 ... 5.0

 °C 

or 

-9.0

 °F .

.. 9.0

 °F 

The zero point of the temperature measuring is shifted for the 
entered value. This can be used to compensate sensor and 
instrument deviations 

** 

 

OFF

 

No zero adjustment for temperature measurement (=0.0°)

 

S[AL 

°C bzw. °F

 

-5.00 ... 5.00 

The slope of the temperature measurement is corrected by this 
value. 
This can be used to compensate sensor and instrument 
deviations 

** 

 

OFF

 

No slope adjustment for temperature measurement (=0.00) 

0FFS 

hPa

 

-20 ... 20

 hPa

 

The zero point of the pressure measuring is shifted for the 
entered value.  
This can be used to compensate sensor deviations 

** 

 

OFF

 

No zero adjustment for pressure measurement (=0 hPa)

 

SET 

AL

 

Set Alarm: Configuration of the alarm settings 

 

13 

AL.  1 

ON    

/

 NO.SO 

Monitoring oxygen: Alarm on with buzzer / Alarm on without 
buzzer 

 

 

OFF 

No alarm monitoring for oxygen 

AL.,N 

[ONC 

Monitoring oxygen: Concentration in % Vol O

2

 

 

 

P. 02

 

Monitoring oxygen: Partial pressure in hPa or mmHg

 

A 1.LO

 

e.g.  

0.0...100.0

 

%

 

Min alarm limit oxygen (not if AL. 1. oFF) 

 

 

A 1.H,

 

e.g.  

0.0...100.0

 

%

 

Max alarm limit oxygen (not if AL. 1. oFF)

 

 

 

AL.  2 

ON    

/

 NO.SO 

Temperature monitoring : Alarm on with buzzer / Alarm on 
without buzzer 

 

 

OFF

 

No alarm monitoring for temperature 

A2.LO

 

-5.0 ...  

+

50.0

 °C 

Min alarm limit temperature (not if AL. 2. oFF) 

 

 

A2.H,

 

-5.0 ... 

+

50.0

 °C 

Max alarm limit temperature (not if AL. 2. oFF) 

 

 

SET 

LO66

  

Set Logger: Configuration of the logger functions 

 

10.3 

FVNC

 

(Y(L

  

Cyclic

: Automatic logger function 

 

STOR

 

Store

: manual recording 

OFF

 

No logger activated 

(Y(L

 

0:0 1 ... 60:00

 

Cycle time in [minutes: seconds] for cyclic logger 

** 

 

SET 

(L0(

 

Set Clock: Configuration of the real time 

 

14 

(L0(

 

HH:MM 

Clock

: set time 

hours: minutes 

 

 

YEAR

 

YYYY 

Year

: set year 

 

 

DATE

 

TT.MM 

Date

: set date 

day: month 

 

 

 

(*)  

If logger memory contains data sets parameters marked with (*) cannot be called. You 

have to clear memory to change these parameters! 
(**)  

If logger is running parameters marked with (**) cannot be called

 

Содержание ResOx 5695

Страница 1: ...ger de www greisinger de Operating manual Water Proof oxygen meter for oxygen in gases with Integrated temperature and pressure measuring and data logger As of version 1 0 ResOx 5695 Please carefully...

Страница 2: ...ENTS 11 7 OXYGEN MEASURING IN GASES PLEASE NOTE 11 7 1 APPLICATION OF THE DIFFERENT SENSOR TYPES 11 8 EXCHANGING THE SENSOR ELEMENT 12 9 CALIBRATION OF THE OXYGEN SENSOR 13 9 1 ONE POINT CALIBRATION A...

Страница 3: ...centration in Vol O2 For the measuring an external sensor of the type GOO GGO or GGA is necessary The measuring take place at the extern opening Sensor Due to the properties of the sensor it has to be...

Страница 4: ...ed to when using the device Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under 20 Speci...

Страница 5: ...s operation When connecting a mains cable or USB interface cable please take care to connect only allowed components The output voltage of a connected power supply unit has to be between 4 5 and 5 5 V...

Страница 6: ...ration Confirm settings return to measuring Store Enter Measuring with Auto Hold off Hold and save current measuring value HLD is displayed with Auto Hold on Start new measuring It is finished when HL...

Страница 7: ...n to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or atta...

Страница 8: ...l configuration 9 12 H 2 P 02 hPa Oxygen partial pressure display in hPa P 02 mmHg Oxygen partial pressure display in mmHg RES K Best O2 resolution LO Low O2 resolution calm value display standard LCD...

Страница 9: ...pressure measuring is shifted for the entered value This can be used to compensate sensor deviations OFF No zero adjustment for pressure measurement 0 hPa SET AL Set Alarm Configuration of the alarm...

Страница 10: ...The sensor state evaluation will be stored after a successful calibration of the oxygen sensor 6 1 2 Mounting operation position The optimum position for the sensor membrane is to point downwards If...

Страница 11: ...have to be flushed with clean air or nitrogen Integrated in GGO 581 GGA 581 GOO 581 Recommended application areas 0 100 Vol O2 Preferred of measurements with extremely small oxygen content e g 0 3 Vol...

Страница 12: ...ynamic pressure due to gas flow this type is optimal It can be screwed tightly into systems with small over or under pressure Attention Mind the maximum pressure The instrument pressure port is connec...

Страница 13: ...e which should be applied 0 2 for pure oxygen NULL for 0 oxygen e g pure nitrogen As long as the display blinks the instrument recognises no valid reference As soon the values for oxygen and temperatu...

Страница 14: ...ate of the recording For the evaluation of the data the software GSOFT3050 version V3 0 or higher has to be used The software also allows easy configuration and starting of the logger When the logger...

Страница 15: ...lection After that press again automatic recording is starting Whenever a measuring is recorded the display shows St XXXXX XXXXX is the number of the measuring If the logger memory is full the display...

Страница 16: ...ure correct sensor calibration adjustment The device provides the following functions to help with this 12 1 Calibration Adjustment interval INT You can input the interval after which the device remin...

Страница 17: ...es battery life time If the device is used together with interface adapter USB 5100 the device is supplied from the interface device pin assignment 1 external supply 5V 50mA 2 GND 3 TxD RxD 3 3V Logic...

Страница 18: ...f display x x x x 176 read min measuring range x 208 read channel count x x x x 177 read max measuring range x 222 read turn off delay x x x x 178 read measuring range unit x 223 Set turn off delay x...

Страница 19: ...ress of water and dust may be lost Avoid unnecessary opening of the instrument 1 Open the 3 Phillips screws at the backside of the instrument 2 Lay down the still closed instrument so that the display...

Страница 20: ...ay range Check Is the value below the specified measuring range value too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error...

Страница 21: ...ressure Lo 0 1100 hPa O2 Hi 0 0 1100 0 hPa O2 Sensor temperature 5 0 50 0 C NTC 10k integrated in GGO GOO GGA cable Absolute pressure 300 5000 hPa abs integrated pressure sensor with pressure port Acc...

Страница 22: ...nction Device will be automatically switched off if no key is pressed no interface communication takes place for the time of the power off delay The power off delay can be set to values between 1and 1...

Отзывы: