H84.0.15.6C-01
Operating manual ResOx 5695
Page 8 of 22
_____________________________________________________
_____________________________________________________________________________
5
Configuration
Some menu points depend on current device settings (e.g. some points are locked if
logger memory contains data sets).
To change
device’s settings, press
for 2 seconds. This will activate the configuration menu (main display
“
SET
“). Pressing
changes between the menus points, pressing
jumps to the referring parameters,
which can be selected with
).
The parameter value can be changed with
or.
. Pressing
again jumps back to the main
configuration menu and saves the settings. Pressing
finishes the configuration.
Pressing
and
at the same time for more than 2 seconds will reset the device to
factory defaults.
If there are data sets stored
(Logger: ”
FVNC STOR
“), the first menu point displayed is
”
READ LO66
“ see chapter 10 Data logger.
If no key is pressed for more than 2 minutes the configuration will be aborted. All changes
will be discarded!
Menu
Parameter Value
Description
or
see
SET
(ONF
Set Configuration: General configuration
9
12
[H 2
P 02
hPa
Oxygen partial pressure display in hPa
*
P 02
mmHg
Oxygen partial pressure display in mmHg
RES
K,
Best O
2
resolution
LO
Low O
2
resolution, calm value display (standard)
LCD.2
T
Secondary display always temperature
P
Secondary display always absolute pressure
P T
Secondary display alternates between temperature and abs.
pressure
U N,T
T
°C
All temperatures in degree Celsius (ex works setting)
°F
All temperatures in degree Fahrenheit
[AL.P
1-PT
Simple one point calibration at air
2-PT
/
3-PT
2 or 3- calibration at air, or in oxygen or in nitrogen / zero gas
[. INT
1
…
365
Calibration reminder period (in days)
OFF
No calibration reminder
A VTO
HLD
ON
Auto stable value determining freezing (when logger = off)
OFF
Standard hold function on key press (when logger = off)
P.OFF
1
…
120
Power-off delay in minutes. Device will be automatically switched
off as soon as this time has elapsed if no key is pressed/no
interface communication takes place. (ex works setting 20min)
OFF
Automatic power-off function deactivated (continuous operation)
L,TE
OFF
Backlight Illumination deactivated
5...120
Turn off backlight
after 5… 120 s (factory settings: 10 s)
ON
Backlight always on
0V T
OFF
Interface off -> minimal power consumption
SER
Serial interface activated (ex works setting)
DA[
Analog output activated