Calibration of the oxygen sensor 

 

In order to compensate for ageing of the sensor, the sensor has to be calibrated at regular 
intervals. 
The device is equipped with an easy-to-use calibration function. 
We recommend to calibrate the sensor at least all 7 days, or to get maximum precision, 
before each measuring series. 

9.1  One point calibration 

(AL.  

 

1-PT

 

The calibration adjusts the sensor to the oxygen content of the atmosphere (20.95 % Vol. O

2

) Therefore 

simply expose the sensor to the ambient air (sufficient ventilation in closed rooms has to be ensured). 

Start calibration: press 

 

for 2 seconds. 

The display will show 

A,R

 

, and as soon the values for oxygen and temperature are stable, the 

calibration will be finished.  
Then the sensor state resulting of the successful calibration will be shown for a short time evaluation in 10% 

steps 

xx%

 

ELE[

9.2  2 / 3- Point calibration 

 (AL 

  

2-PT

(AL 

  

3-PT

 

The sensor will be automatically calibrated to the oxygen content of the atmosphere (20.95 % Vol. O

2

), and 

one or two additional concentrations. As reference, gases usually Nitrogen (0 % Vol. O

2

) or pure oxygen are 

used (100 % Vol. O

2

). 

1.  Start calibration: press: 

 

- key for 2 seconds. 

2.  First calibration reference: (Pt.1) 

As first reference at a 3-point calibration, the zero reference has to be applied 0 % VOL O

2

at a 2-point calibration either 100 % Vol. O

2

 or 0 % Vol. O

2.

 

The display will show 

, and the referring reference which should be applied. 

0.2

 for pure oxygen. 

NULL

 for 0% oxygen (e. g. pure nitrogen). 

As long as the display blinks, the instrument recognises no valid reference. 
As soon the values for oxygen and temperature are stable, the calibration of the first point will be finished

.

 

The instrument tells you to apply the next reference (possible references are blinking in the display). 

3.  Second calibration reference: (Pt.2) 

The display will show 

, and the referring reference which should be applied. 

A,R

 for ambient air. 

0.2

 for pure oxygen. 

NULL

 for 0 % oxygen. (e. g. pure nitrogen). 

As long as the display blinks, no valid reference is recognised by the instrument. 
As soon the values for oxygen and temperature are stable, the calibration of the second point will be 
finished. At 2-point calibration the calibration will be finished and the sensor state resulting of the successful 

calibration will be shown for a short time evaluation in 10% steps 

xx%

 

ELE[

At 3-point calibration, the instrument tells you to apply the next reference (possible reference is blinking). 

1.  Third calibration reference: (Pt.3) 

The display will show 

, and the referring reference which should be applied: 

As soon the values for oxygen and temperature are stable, the calibration of the second point will be 
finished. At 2-point calibration the calibration will be finished and the sensor state resulting of the successful 

calibration will be shown for a short time evaluation in 10% steps 

xx%

 

ELE[

 

In case of error messages being displayed during the calibration process, please refer to 
our notes at the end of this manual! If a calibration cannot be carried out after an extended 
period of time, at least one of the measuring values is unstable (oxygen partial pressure, 
temperature).  
Please check your measuring arrangements! 

 

 

Содержание ResOx 5695

Страница 1: ...ger de www greisinger de Operating manual Water Proof oxygen meter for oxygen in gases with Integrated temperature and pressure measuring and data logger As of version 1 0 ResOx 5695 Please carefully...

Страница 2: ...ENTS 11 7 OXYGEN MEASURING IN GASES PLEASE NOTE 11 7 1 APPLICATION OF THE DIFFERENT SENSOR TYPES 11 8 EXCHANGING THE SENSOR ELEMENT 12 9 CALIBRATION OF THE OXYGEN SENSOR 13 9 1 ONE POINT CALIBRATION A...

Страница 3: ...centration in Vol O2 For the measuring an external sensor of the type GOO GGO or GGA is necessary The measuring take place at the extern opening Sensor Due to the properties of the sensor it has to be...

Страница 4: ...ed to when using the device Trouble free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under 20 Speci...

Страница 5: ...s operation When connecting a mains cable or USB interface cable please take care to connect only allowed components The output voltage of a connected power supply unit has to be between 4 5 and 5 5 V...

Страница 6: ...ration Confirm settings return to measuring Store Enter Measuring with Auto Hold off Hold and save current measuring value HLD is displayed with Auto Hold on Start new measuring It is finished when HL...

Страница 7: ...n to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or atta...

Страница 8: ...l configuration 9 12 H 2 P 02 hPa Oxygen partial pressure display in hPa P 02 mmHg Oxygen partial pressure display in mmHg RES K Best O2 resolution LO Low O2 resolution calm value display standard LCD...

Страница 9: ...pressure measuring is shifted for the entered value This can be used to compensate sensor deviations OFF No zero adjustment for pressure measurement 0 hPa SET AL Set Alarm Configuration of the alarm...

Страница 10: ...The sensor state evaluation will be stored after a successful calibration of the oxygen sensor 6 1 2 Mounting operation position The optimum position for the sensor membrane is to point downwards If...

Страница 11: ...have to be flushed with clean air or nitrogen Integrated in GGO 581 GGA 581 GOO 581 Recommended application areas 0 100 Vol O2 Preferred of measurements with extremely small oxygen content e g 0 3 Vol...

Страница 12: ...ynamic pressure due to gas flow this type is optimal It can be screwed tightly into systems with small over or under pressure Attention Mind the maximum pressure The instrument pressure port is connec...

Страница 13: ...e which should be applied 0 2 for pure oxygen NULL for 0 oxygen e g pure nitrogen As long as the display blinks the instrument recognises no valid reference As soon the values for oxygen and temperatu...

Страница 14: ...ate of the recording For the evaluation of the data the software GSOFT3050 version V3 0 or higher has to be used The software also allows easy configuration and starting of the logger When the logger...

Страница 15: ...lection After that press again automatic recording is starting Whenever a measuring is recorded the display shows St XXXXX XXXXX is the number of the measuring If the logger memory is full the display...

Страница 16: ...ure correct sensor calibration adjustment The device provides the following functions to help with this 12 1 Calibration Adjustment interval INT You can input the interval after which the device remin...

Страница 17: ...es battery life time If the device is used together with interface adapter USB 5100 the device is supplied from the interface device pin assignment 1 external supply 5V 50mA 2 GND 3 TxD RxD 3 3V Logic...

Страница 18: ...f display x x x x 176 read min measuring range x 208 read channel count x x x x 177 read max measuring range x 222 read turn off delay x x x x 178 read measuring range unit x 223 Set turn off delay x...

Страница 19: ...ress of water and dust may be lost Avoid unnecessary opening of the instrument 1 Open the 3 Phillips screws at the backside of the instrument 2 Lay down the still closed instrument so that the display...

Страница 20: ...ay range Check Is the value below the specified measuring range value too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error...

Страница 21: ...ressure Lo 0 1100 hPa O2 Hi 0 0 1100 0 hPa O2 Sensor temperature 5 0 50 0 C NTC 10k integrated in GGO GOO GGA cable Absolute pressure 300 5000 hPa abs integrated pressure sensor with pressure port Acc...

Страница 22: ...nction Device will be automatically switched off if no key is pressed no interface communication takes place for the time of the power off delay The power off delay can be set to values between 1and 1...

Отзывы: