 Operating manual GMH 5430 

page 10 of 18 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

GHM GROUP 

– Greisinger  |  GHM Messtechnik GmbH 

Hans-Sachs-Str. 26  |  93128 Regenstauf  |  GERMANY  
Phone +49 9402 9383-0  |  Fax +49 9402 9383-33 

www.greisinger.de

  |  

[email protected]

e

 

LF

Tref

=

LF

Tx

 1 +

TK

lin

100% • (Tx − Tref)

 

Temperature coefficient of about 2.0 %/K are most common.  
A temperature coefficient can be determined for example by measuring a solution with deactivated 
temperature compensation at two different temperatures (T1 and T2). 

TK

lin

=

(LF

T1

− LF

T2

) • 100%

(T1 − T2) • LF

T1

 

TK

lin

 

is the value input at the menu “t.Lin”. 

LF

T1

 

conductivity at temperature T1 

LF

T2

 conductivity at temperature T2  

7  Configuration 

 

Some menu points depend on current device settings. 

 

To change d

evice’s settings, press 

“menu” 

 

for 2 seconds. This will activate the configuration menu 

(main display: “SEt”). Pressing 

“menu” 

 

changes between the menus points, pressing 

 jumps to the 

referring parameters, which can be selected with key 

. The parameter value can be changed with 

 or 

. Pressing 

“menu” 

 

again jumps back to the main configuration menu and saves the settings.  

Pressing

 

“enter” 

 

finishes the configuration.

 

 

Pressing “menu” and “store” at the same time for more than 2 seconds will reset the device to 
factory defaults. 

If no key is pressed for more than 2 minutes the configuration will be aborted. All changes will be 
discarded! 

Menu 

Parameter 

Value 

Description

 

 

 

 

 

 or 

 

 

 

 

 

 

Set Configuration: General configuration

 

 

 

 

I

nput: Selection of measured unit

 

** 

 

SEt 

Choice of displayed unit via set-key 

Cond 

Conductivity 

rES

 

Resistivity 

SAL

 

Salinity 

tdS

 

Total dissolved solids 

 

TDS measurement: conversion factor (only if Inp = tdS) 

 

 

0.40 - 1.00 

Conversion factor for TDS measurement 

 

Cell Range: Adjustment of cell constant: cell constant range 

 

 

0.01 

Ultrapure water, electrodes with K ~ 0.01 

0.1 

Ultrapure water, electrodes with K ~ 0.1 

Standard electrodes, i.e. with K= 0.55 or K=1 

10 

Electrodes with K=10 

 

Cell Factor: Adjustment of cell constant: multiplication factor  

 

 

0.4000 -   
 1.5000

 

Multiplication factor of cell constant  
Cell constant CELL = CELL Range * CELL Factor 

 

t-Input: Selection of temperature input type 

 

 

 

ntc 

NTC 10k sensor 

Pt 

Pt1000 sensor 

 

Range: Selection of display range (conductivity, resistivity or tdS)  

 

 

Auto 

Automatic range selection 

e.g. 0.0 - 500.0 µS/cm  Example for CELL rAng 1 and InP Cond: others in chapter 6.2 
…0- 1000 mS/cm 

Example for CELL rAng 1 and InP Cond: others in chapter 6.2 

 

 

Содержание GMH 5430

Страница 1: ...2 6C 10 Operating Manual Conductivity measuring device waterproof GMH 5430 Keep for later use www ghm group de Companies Brands of the GHM Englisch...

Страница 2: ...TY 7 6 2 MEASURING RANGES AND CELL CONSTANTS 7 6 3 CONDUCTIVITY MEASUREMENT 7 6 4 RESISTIVITY MEASUREMENT 8 6 5 TDS MEASUREMENT 8 6 6 SALINITY MEASUREMENT 8 6 7 ELECTRODES MEASURING CELLS 9 6 7 1 Assi...

Страница 3: ...have further knowledge in case of doubt the user is liable for any harm damage resulting from misinterpretation due to insufficient knowledge The manufacturer will assume no liability or warranty in...

Страница 4: ...in personal injury or material damage Failure to comply with these instructions could result in death or serious injury and material damage 5 This device must not be used at potentially explosive area...

Страница 5: ...ain display conductivity mS cm S cm resistivity k cm M cm TDS total dissolved solids mg l salinity SAL 2 Secondary display measuring value temperature 3 Arrows to selected measuring unit 4 Rating of b...

Страница 6: ...se of bayonet connectors Lock unlock with turnable ring at cable socket Protect contacts from moisture and soiling 4 4 Pop up clip Handling Pull at label open in order to swing open the pop up clip Pu...

Страница 7: ...ng on the used electrode Therefore the device offers four cell constant ranges to choose depending on the correspondent cell constant K CELL rAnG Selectable cell constant K Application 0 01 0 004000 0...

Страница 8: ...types NaCl KCl etc 0 50 Natural waters surface waters drinking water 0 65 0 70 e g salt concentration of aqueous fertilizer solutions Attention This are only approximate values good for estimations b...

Страница 9: ...f measuring cells Especially in harsh environments and over long time the cell constants of measuring cells are drifting Depending on the application and use we recommend a regular checking of the pre...

Страница 10: ...main configuration menu and saves the settings Pressing enter finishes the configuration Pressing menu and store at the same time for more than 2 seconds will reset the device to factory defaults If...

Страница 11: ...Reference temperature 25 C 77 F 20 C 68 F Reference temperature 20 C 68 F Adjustment Calibration Adjustment reminder period factory setting 180 1 730 Adjustment reminder period in days oFF No adjustm...

Страница 12: ...want to develop your own software we offer a GMH3000 development package including a universally applicable Windows functions library GMH3000 DLL with documentation can be used by all established pro...

Страница 13: ...with enter Afterwards the device returns to the normal measuring operation mode or if so displays an error message The resulting cell constant can be seen in the menu at CELL rAng and the calibration...

Страница 14: ...ead out via PC software GMHKonfig or displayed directly at the device Press for 2 seconds The display will show configuration level Press several times until this is displayed read cal read calibratio...

Страница 15: ...g taking care that it is positioned correctly otherwise the sealing may be damaged Afterwards press the two halves together lay the instrument with display pointing downwards and screw it together aga...

Страница 16: ...0 C at 111 8 mS cm If bAt is flashing the battery will be exhausted soon Further measurements are possible for short time If bAt is displayed continuously the battery is ultimately exhausted and has t...

Страница 17: ...ifferent measuring cells Supported temperature sensors Pt1000 and NTC 10k Interface ext supply 4 pole bayonet socket for serial interface and supply USB Adapter USB 5100 Display 4 digit 7 segment illu...

Страница 18: ..._________________________________ _____________________________________________________________________________ GHM GROUP Greisinger GHM Messtechnik GmbH Hans Sachs Str 26 93128 Regenstauf GERMANY Pho...

Отзывы: