1  General Note 

Read this document carefully and get used to the operation of the device before you use it. Keep this 
document within easy reach near the device for consulting in case of doubt. 

2  Safety 

2.1  Intended Use 

The device is designed for measuring conductivity, resistivity, salinity and TDS in fluids 

– using external 

suitable electrodes (measuring cells). It is designed for the mobile use or the stationary operation in a 
controlled electromagnetic environment (lab). The electrodes are connected via 7-pole bayonet connection. 

Please consider:  Depending on the measuring range different electrode types may be needed 

– choose an 

appropriate one. 

Personnel which starts up, operates and maintains the device has to have sufficient knowledge of the 
measuring procedure and the meaning of the resulting measured values, this manual delivers a valuable 
help for this. The instructions of the manual have to be understood, regarded and followed.  

To be sure that there´s no risk arising due to misinterpretation of measured values, the operator must have 
further knowledge in case of doubt - the user is liable for any harm/damage resulting from misinterpretation 
due to insufficient knowledge. 

The manufacturer will assume no liability or warranty in case of usage for other purpose than the intended 
one, ignoring this manual, operating by unqualified staff as well as unauthorized modifications to the device. 

2.2  Safety signs and symbols 

Warnings are labeled in this document with the followings signs: 
 

 

Caution! 

This symbol warns of imminent danger, death, serious injuries 

and significant damage to property at non-observance. 

 

Attention! 

This symbol warns of possible dangers or dangerous 

situations which can provoke damage to the device or environment at 
non-observance. 

 

Note! 

This symbol point out processes which can indirectly influence 

operation or provoke unforeseen reactions at non-observance.

 

2.3  Safety guidelines 

This device has been designed and tested in accordance with the safety regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures 
and special safety advises given in this manual will be adhered to when using the device. 

1.  Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to 

any other climatic conditions than those stated under "Specification". 
If the device is transported from a cold to a warm environment condensation may cause in a failure of the 
function. In such a case make sure the device temperature has adjusted to the ambient temperature 
before trying a new start-up. 

 

 

Содержание GMH 5430

Страница 1: ...2 6C 10 Operating Manual Conductivity measuring device waterproof GMH 5430 Keep for later use www ghm group de Companies Brands of the GHM Englisch...

Страница 2: ...TY 7 6 2 MEASURING RANGES AND CELL CONSTANTS 7 6 3 CONDUCTIVITY MEASUREMENT 7 6 4 RESISTIVITY MEASUREMENT 8 6 5 TDS MEASUREMENT 8 6 6 SALINITY MEASUREMENT 8 6 7 ELECTRODES MEASURING CELLS 9 6 7 1 Assi...

Страница 3: ...have further knowledge in case of doubt the user is liable for any harm damage resulting from misinterpretation due to insufficient knowledge The manufacturer will assume no liability or warranty in...

Страница 4: ...in personal injury or material damage Failure to comply with these instructions could result in death or serious injury and material damage 5 This device must not be used at potentially explosive area...

Страница 5: ...ain display conductivity mS cm S cm resistivity k cm M cm TDS total dissolved solids mg l salinity SAL 2 Secondary display measuring value temperature 3 Arrows to selected measuring unit 4 Rating of b...

Страница 6: ...se of bayonet connectors Lock unlock with turnable ring at cable socket Protect contacts from moisture and soiling 4 4 Pop up clip Handling Pull at label open in order to swing open the pop up clip Pu...

Страница 7: ...ng on the used electrode Therefore the device offers four cell constant ranges to choose depending on the correspondent cell constant K CELL rAnG Selectable cell constant K Application 0 01 0 004000 0...

Страница 8: ...types NaCl KCl etc 0 50 Natural waters surface waters drinking water 0 65 0 70 e g salt concentration of aqueous fertilizer solutions Attention This are only approximate values good for estimations b...

Страница 9: ...f measuring cells Especially in harsh environments and over long time the cell constants of measuring cells are drifting Depending on the application and use we recommend a regular checking of the pre...

Страница 10: ...main configuration menu and saves the settings Pressing enter finishes the configuration Pressing menu and store at the same time for more than 2 seconds will reset the device to factory defaults If...

Страница 11: ...Reference temperature 25 C 77 F 20 C 68 F Reference temperature 20 C 68 F Adjustment Calibration Adjustment reminder period factory setting 180 1 730 Adjustment reminder period in days oFF No adjustm...

Страница 12: ...want to develop your own software we offer a GMH3000 development package including a universally applicable Windows functions library GMH3000 DLL with documentation can be used by all established pro...

Страница 13: ...with enter Afterwards the device returns to the normal measuring operation mode or if so displays an error message The resulting cell constant can be seen in the menu at CELL rAng and the calibration...

Страница 14: ...ead out via PC software GMHKonfig or displayed directly at the device Press for 2 seconds The display will show configuration level Press several times until this is displayed read cal read calibratio...

Страница 15: ...g taking care that it is positioned correctly otherwise the sealing may be damaged Afterwards press the two halves together lay the instrument with display pointing downwards and screw it together aga...

Страница 16: ...0 C at 111 8 mS cm If bAt is flashing the battery will be exhausted soon Further measurements are possible for short time If bAt is displayed continuously the battery is ultimately exhausted and has t...

Страница 17: ...ifferent measuring cells Supported temperature sensors Pt1000 and NTC 10k Interface ext supply 4 pole bayonet socket for serial interface and supply USB Adapter USB 5100 Display 4 digit 7 segment illu...

Страница 18: ..._________________________________ _____________________________________________________________________________ GHM GROUP Greisinger GHM Messtechnik GmbH Hans Sachs Str 26 93128 Regenstauf GERMANY Pho...

Отзывы: