Operating manual GMH 5430
page 16 of 18
_____________________________________________________
_____________________________________________________________________________
GHM GROUP
– Greisinger | GHM Messtechnik GmbH
Hans-Sachs-Str. 26 | 93128 Regenstauf | GERMANY
Phone +49 9402 9383-0 | Fax +49 9402 9383-33
www.greisinger.de
|
e
upper display
expired or last calibration is not valid
Error messages for automatic cell constant adjustment/calibration:
CAL Err.1
Cell constant too high
Determined cell constant must not exceed
1.2* cell range
CAL Err.2
Cell constant too small
Determined cell constant must not fall below
0.4* cell range
CAL Err.3
Solution of wrong range
Wrong cell range / wrong solution / far
beyond tolerance
CAL Err.4
Wrong temperature
Beyond permitted temperature:
0.0
– 34.0 °C
(or 0.0
– 27.0 °C at 111.8 mS/cm)
If
“bAt”
is flashing the battery will be exhausted soon. Further measurements are possible for short time. If
“bAt” is displayed continuously the battery is ultimately exhausted and has to be replaced. Further
measurements aren’t possible any more.
15 Reshipment and Disposal
15.1 Reshipment
All devices returned to the manufacturer have to be free of any residual of measuring media
and other hazardous substances. Measuring residuals at housing or sensor may be a risk for
persons or environment
Use an adequate transport package for reshipment, especially for fully functional devices.
Please make sure that the device is protected in the package by enough packing materials.
15.2 Disposal instructions
Batteries must not be disposed in the regular domestic waste but at the designated collecting
points.
The device must not be disposed in the unsorted municipal waste! Send the device directly to us
(sufficiently stamped), if it should be disposed. We will dispose the device appropriate and
environmentally sound.