13 Replacing Batteries
Before changing batteries, please read the following instruction and
follow it step by step.
Not following the instruction may cause harm to the instrument or the
protection against ingress of water and dust may be lost!
Avoid unnecessary opening of the instrument!
1. Open the 3 Phillips screws at the backside of the instrument.
2. Lay down the still closed instrument, so that the display side
points upwards.
The lower half of the housing incl. the electronics should be kept
lying down during battery change.
This avoids loss of the 3 sealing rings placed in the screw holes.
3. Lift off upper half of housing. Keep an eye on the six function
keys, to be sure not to damage them.
4. Change carefully the two batteries (Type: AAA).
5. Check: Are the 3 sealing rings placed in the housing?
Is the circumference seal of the upper half sound and clean?
6.
Close the housing, taking care that it is positioned correctly,
otherwise the sealing may be damaged. Afterwards press the
two halves together, lay the instrument with display pointing
downwards and screw it together again
Take care to screw only until you feel increasing resistance,
higher screwing force does not result in higher water
protection!
14 Error and System Messages
Error messages for measurement
Description
What to do?
No display or
confused
characters,
Device does not
react on keypress
Battery empty
Replace battery
Mains operation: wrong voltage or polarity
Check power supply, replace it if necessary
System error
Disconnect battery and power supply, wait shortly,
then reconnect
Device defective
Return to manufacturer for repair
Err.1
Measured value above allowable range
Check: pressure not within sensor range?
-> measuring value to high!
Sensor defective
Return to manufacturer for repair
Err.2
Measured value below allowable range
Check: pressure not within sensor range?
-> measuring value to low!
Sensor defective
Return to manufacturer for repair
Err.7
System error
Return to manufacturer for repair
Value extremely out of measuring range
Value extremely out of measuring range
----
Could not calculate display value
• measuring range or input range
exceeded
Check range parameter
• measured values are instable
Wait for signal regulation of the device
> CAL <
CAL flashing in
Either preset calibration interval has
Device has to be calibrated!