H83.0.02.6C-10 

 Operating manual GMH 5430 

page 4 of 18 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

GHM GROUP 

– Greisinger  |  GHM Messtechnik GmbH 

Hans-Sachs-Str. 26  |  93128 Regenstauf  |  GERMANY  
Phone +49 9402 9383-0  |  Fax +49 9402 9383-33 

www.greisinger.de

  |  

[email protected]

e

 

2.    

 

If there is a risk whatsoever involved in running it, the device has to be switched off 
immediately and to be marked accordingly to avoid re-starting. 
Operator safety may be a risk if: 
- there is visible damage to the device 
- the device is not working as specified 
- the device has been stored under unsuitable conditions for a longer time. 
In case of doubt, please return device to manufacturer for repair or maintenance. 

3.  When connecting the device to other devices the connection has to be designed most thoroughly as 

internal connections in third-party devices (e.g. connection GND with protective earth) may lead to 
undesired voltage potentials that can lead to malfunctions or destroying of the instrument and the 
connected devices. 
 

This device must not be run with a defective or damaged power supply unit.  
Danger to life due to electrical shock!  
 

 

4. 

 

 

Do not use these products as safety or emergency stop devices or in any other application 
where failure of the product could result in personal injury or material damage. 
Failure to comply with these instructions could result in death or serious injury and material 
damage. 

 

5. 

 

 

This device must not be used at potentially explosive areas! The usage of this device at 
potentially explosive areas increases danger of deflagration, explosion or fire due to 
sparking. 

Product Specification 

3.1  Scope of supply 

The scope of supply includes:  

  GMH 5430 with 2 AAA batteries 

  Operating manual 

  Short form manual 

3.2  Operation and maintenance advice 

1.  Temperature measuring / sensor connection: 

There is the possibility to connect a temperature sensor (Pt1000 or NTC 10k) to the 7-pole bayonet 
socket. Generally a suitable temperature sensor is included to the measuring cell. The measured 
temperature is used for the automatic temperature compensation (e.g. Lin or nlF) and is additionally 
displayed. 

2.  Battery operation: 

If ‘bAt’ is shown in the lower display the battery has been used up and needs to be replaced. However, 
the device will operate correctly for a certain time. If ‘bAt’ is shown in the upper display the voltage is too 
low to operate the device; the battery has been completely used up. Battery change: p.r.t. chapter 13. 

The battery has to be taken out, when storing device above 50°C. We recommend taking out 
battery if device is not used for a longer period of time. 
After recommissioning the real-time clock has to be set again.  

3.  Treat device and sensor carefully. Use only in accordance with above specification. (do not throw, hit 

against etc.). Protect plug and socket from soiling. 

 

 

Содержание GMH 5430

Страница 1: ...2 6C 10 Operating Manual Conductivity measuring device waterproof GMH 5430 Keep for later use www ghm group de Companies Brands of the GHM Englisch...

Страница 2: ...TY 7 6 2 MEASURING RANGES AND CELL CONSTANTS 7 6 3 CONDUCTIVITY MEASUREMENT 7 6 4 RESISTIVITY MEASUREMENT 8 6 5 TDS MEASUREMENT 8 6 6 SALINITY MEASUREMENT 8 6 7 ELECTRODES MEASURING CELLS 9 6 7 1 Assi...

Страница 3: ...have further knowledge in case of doubt the user is liable for any harm damage resulting from misinterpretation due to insufficient knowledge The manufacturer will assume no liability or warranty in...

Страница 4: ...in personal injury or material damage Failure to comply with these instructions could result in death or serious injury and material damage 5 This device must not be used at potentially explosive area...

Страница 5: ...ain display conductivity mS cm S cm resistivity k cm M cm TDS total dissolved solids mg l salinity SAL 2 Secondary display measuring value temperature 3 Arrows to selected measuring unit 4 Rating of b...

Страница 6: ...se of bayonet connectors Lock unlock with turnable ring at cable socket Protect contacts from moisture and soiling 4 4 Pop up clip Handling Pull at label open in order to swing open the pop up clip Pu...

Страница 7: ...ng on the used electrode Therefore the device offers four cell constant ranges to choose depending on the correspondent cell constant K CELL rAnG Selectable cell constant K Application 0 01 0 004000 0...

Страница 8: ...types NaCl KCl etc 0 50 Natural waters surface waters drinking water 0 65 0 70 e g salt concentration of aqueous fertilizer solutions Attention This are only approximate values good for estimations b...

Страница 9: ...f measuring cells Especially in harsh environments and over long time the cell constants of measuring cells are drifting Depending on the application and use we recommend a regular checking of the pre...

Страница 10: ...main configuration menu and saves the settings Pressing enter finishes the configuration Pressing menu and store at the same time for more than 2 seconds will reset the device to factory defaults If...

Страница 11: ...Reference temperature 25 C 77 F 20 C 68 F Reference temperature 20 C 68 F Adjustment Calibration Adjustment reminder period factory setting 180 1 730 Adjustment reminder period in days oFF No adjustm...

Страница 12: ...want to develop your own software we offer a GMH3000 development package including a universally applicable Windows functions library GMH3000 DLL with documentation can be used by all established pro...

Страница 13: ...with enter Afterwards the device returns to the normal measuring operation mode or if so displays an error message The resulting cell constant can be seen in the menu at CELL rAng and the calibration...

Страница 14: ...ead out via PC software GMHKonfig or displayed directly at the device Press for 2 seconds The display will show configuration level Press several times until this is displayed read cal read calibratio...

Страница 15: ...g taking care that it is positioned correctly otherwise the sealing may be damaged Afterwards press the two halves together lay the instrument with display pointing downwards and screw it together aga...

Страница 16: ...0 C at 111 8 mS cm If bAt is flashing the battery will be exhausted soon Further measurements are possible for short time If bAt is displayed continuously the battery is ultimately exhausted and has t...

Страница 17: ...ifferent measuring cells Supported temperature sensors Pt1000 and NTC 10k Interface ext supply 4 pole bayonet socket for serial interface and supply USB Adapter USB 5100 Display 4 digit 7 segment illu...

Страница 18: ..._________________________________ _____________________________________________________________________________ GHM GROUP Greisinger GHM Messtechnik GmbH Hans Sachs Str 26 93128 Regenstauf GERMANY Pho...

Отзывы: