3. Diagnose:
When LCD indicates main menu, press button SELECTION to move the
diagnosis of the arrow, LCD indicates as follows:
BASIC SETTING
INDICATION
DIAGNOSE
SERVICE
INFORMATION
The function of diagnose can be used for testing the working conditions of
the instrument and other parts and debugging the system. Press button
OK to enter the function of diagnose.
3.1. Peak
measurement:
Peak indication is the indication of the distance peak during the
measurement. This parameter can be deleted with the reset under item
4.4 of service menu. LCD indicates as follows:
MEASUREMENT PEAK 3.1
MINIMUM DISTANCE VALUE
0.000 m (d)
MAXIMUM DISTANCE VALUE 2.109 m (d)
3.2. Measurement
mode:
When LCD indicates the menu number 3.1, press button SELECTION to
enter the next diagnose mode. LCD indicates as follows:
MEASUREMENT MODE 3.2
MEASUREMENT RELIABILITY:
20 dB
SENSOR MODE:
OK
3.3. Curve
selection:
When LCD indicates the menu number 3.2, press button SELECTION to
enter graph indication mode. LCD indicates as follows:
CURVE SELECTION 3.3
ECHO GRAPH
When need to select other curves, press button OK to enter curve
selection menu. LCD indicates as follows:
CURVE SELECTION 3.3
ECHO GRAPH
FALSE ECHO GRAPH
OUTPUT TREND CURVE
Press button SELECTION to move the arrow to the point which is required
to be indicated along the curve, press button OK for confirmation.
3.4. ECHO GRAPH:
When LCD indicates the menu number 3.3, press button SELECTION to
make LCD indicate the selected curve. The function of curve zoom can be
used to amplify the curve along the axes of time and amplitude so that it to
be read clearly. When LCD indicates graph, press button OK to enter
curve zoom editing menu. LCD indicates as follows:
CURVE ZOOM 3.4
ZOOM ALONG AXIS X
ZOOM ALONG AXIX Y
NO ZOOM
Press button SELECTION to move the arrow for selecting zoom direction
or no zoom, press button OK
for confirmation. When select zoom along axis X, press button
SELECTION to move the start point to the place where it is needed, press
button OK for confirmation. Press button SELECTION again to move the
end point to the place where it is needed, press button OK for
confirmation. At the time, the area selected is amplified up to the whole