7.6. End of Freezing Profile
When the freezing profile is complete
the ‘Program Time Remaining’ will show ‘Complete’ on
the Monitor panel. On the freezer unit
the yellow ‘Complete’ light will illuminate and the buzzer
will sound. The samples can now be removed from the freezer and placed in long term
storage. The freezer will hold at the final temperature of the c
ycle until the ‘Stop Cycle’ button
is pressed or the freezer is switched off. Plate temperature data will continue to be logged.
7.7. Stopping the Cycle and Generating a Report.
When all of the vessels have bee
n transferred, click on the ‘Stop Cycle’ button on the
software Monitor panel. Temperature logging will stop and the unit will return to the start
temperature. The Cryocooler Stat
us changes to ‘Temp Stabilization.’
A message box will appear asking if a
report is required. Click ‘Yes’ to generate a report in pdf
or TSV (tab separated
value) format. Click ‘No’ if a report is not required immediately. It is
recommended that a report is generated and checked after each run.
Figure 46: Monitor Panel. End of Cycle
Reports can still be generated at any time after the cycle has been stopped using the Report
Panel.