Tidal Volume:
Range dependent on ventilation settings, accuracy: ±10%
with 100% Oxygen or 100% Air (±12% below -9
o
C), ±12% for blended
gases (±14% below -9
o
C)
High Airway Alarm:
Adjustable 15 – 80 cmH
2
O, accuracy: ±5% or ±1
cmH
2
O, whichever is greater
Safety Pressure Relief:
Fixed at 88 cmH
2
O, accuracy: ±10% cmH
2
O
External Alarm Connection:
1⁄4” Phono style jack, normally open switch
that will close when an alarm occurs. Max 0.5 Amp at 125 VAC or 1 Amp
at 30VDC signal. Minimum signal 1 mAmp @ 5 volts.
WARNING:
Use only insulated connectors for the external alarm
connection. The use of non-insulated connectors will result in a shock
hazard if using high voltages.
Airway Pressure:
0 – 99 cmH
2
O, accuracy: ±5% or ±1 cmH
2
O (whichever
is greater)
Internal Battery Life:
Run time at 21°C (70±5°F), BPM = 10, and Vt =
600 mL
•
100% O
2
: 7.5 hours (Approx 5 hours at 0
°
F (-18°C))
•
100% Air: 7.5 hours (Approx 5 hours at 0
°
F (-18°C))
•
21% (Air Compressor): 7 hours (Approx 4 hours at 0
°
F (-18°C))
Oxygen Inlet Filter:
65 micron sintered bronze
Burst Pressure:
145 psi (1000kPa) minimum through oxygen inlet
Storage Conditions:
-40 – 60°C (-40 – 140°F), 10% - 95% RH non-
condensing
Shipping Conditions:
-40 – 60°C (-40 – 140°F), 5% - 95% RH non-
condensing
Sound Level:
tested per ISO 80601-2-12
•
< 40 dBA with compressed gas
•
< 50 dBA with internal compressor and flows less than 36 LPM
•
< 55 dBA max sound level while using internal compressor
NOTE: Maintenance schedules must be followed to ensure the
product can reliably meet specifications.