22
5)
To store the altered settings press and hold the button of the mode
being edited (Adult or Child) for 3 seconds, until the ventilator
beeps. The ventilator will be operating with the stored settings.
Note that for CPAP and BIPAP, because the CCV is operating, it will go
into Apnea Ventilation Mode in 20 seconds. If the adjustment is taking
longer than that, breathe with the circuit to trigger a spontaneous
breath. In Pressure Control and Volume Control the use of a test lung is
recommended to prevent alarms from activating.
Revising a Custom Mode
1)
Make sure the Control Lock is not activated. When it is on, a blue
light illuminates above it. Control Lock may be used to provide
additional protection against accidentally changing any setting.
2)
Push the Custom Mode Selection button and use the
Parameter/mode Adjustment arrows to move the blinking light to
the custom mode to be revised.
3)
Push the Custom Mode Selection button again to begin using that
mode. The blinking light will become solid.
4)
Push and hold the Custom Mode Selection button for 3 seconds. The
light by the selected custom mode will now flash rapidly.
5)
Revise any settings as desired, using the Parameter/mode
Adjustment arrows.
6)
To store the revised settings, press and hold the Custom Mode
Selection button first and then the Mode Selection button. The
ventilator will sound a short beep to indicate that the Custom Mode
has been saved.